Table of Contents Table of Contents
Previous Page  83 / 629 Next Page
Information
Show Menu
Previous Page 83 / 629 Next Page
Page Background www.restek.com

81

GC COLUMNS |

APPLICATION-SPECIFIC COLUMNS

Environmental

Rxi®-624Sil MS Columns

(fused silica)

(midpolarity Crossbond® phase)

• Low-bleed, high-thermal stability column—maximum temperatures up to 320 °C.

• Inert—excellent peak shape for a wide range of compounds.

• Selective—G43 phase highly selective for volatile organics and residual solvents,

great choice for USP<467>.

• Manufactured for column-to-column reproducibility—well-suited for validated

methods.

Rxi®-624Sil MS (G43) Structure

x

Si

O

(CH

2

)

3

CH

3

CH

3

CH

3

Si

O

R

O

CH

3

CH

3

Si

94%

y

CH

3

CH

3

Si

C N

similar phases

DB-624, VF-624ms, CP-Select 624 CB

20-Meter

30-Meter

60-Meter

75-Meter

105-Meter

ID df

temp. limits cat.#

price

cat.#

price

cat.#

price

cat.#

price

cat.#

price

0.18 mm 1.00 µm -20 to 300/320 °C 13865

0.25 mm 1.40 µm -20 to 300/320 °C

13868

13869

0.32 mm 1.80 µm -20 to 300/320 °C

13870

13872

0.53 mm 3.00 µm -20 to 280/300 °C

13871

13873

13874

13875

Volatiles by EPA Method 8260 on Rxi®-624Sil MS (30 m, 0.25 mm ID, 1.40 µm)

GC_EV1169

Column

Rxi®-624Sil MS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 13868)

Sample

8260A surrogate mix (cat.# 30240)

8260A internal standard mix (cat.# 30241)

8260B MegaMix® calibration mix (cat.# 30633)

VOA calibration mix #1 (ketones) (cat.# 30006)

8260B acetate mix (Revised) (cat.# 30489)

California oxygenates mix (cat.# 30465)

502.2 calibration mix #1 (gases) (cat.# 30042)

Conc.:

25 ppb in RO water

Injection

Purge and trap split (split ratio 30:1)

Inj. Temp.:

225 °C

Purge and Trap

Instrument:

OI Analytical 4660

Trap Type:

10 Trap

Purge:

11 min @ 20 °C

Desorb Preheat Temp.: 180 °C

Desorb:

0.5 min @ 190 °C

Bake:

5 min @ 210 °C

Interface Connection: Injection port

Oven

Oven Temp.:

35 °C (hold 5 min) to 60 °C at 11 °C/min to 220 °C

at 20 °C/min (hold 2 min)

Carrier Gas

He, constant flow

Flow Rate:

1.0 mL/min

Detector

MS

Mode:

Scan

Transfer Line Temp.:

230 °C

Analyzer Type:

Quadrupole

Source Temp.:

230 °C

Quad Temp.:

150 °C

Electron Energy:

70 eV

Solvent Delay Time:

1.5 min

Tune Type:

BFB

Ionization Mode:

EI

Scan Range:

36-260 amu

Instrument

Agilent 7890A GC

& 5975C MSD

Notes

Other Purge-and-Trap Conditions:

Sample Inlet: 40 °C

Sample: 40 °C

Water Management: Purge 110 °C, Desorb 0 °C, Bake, 240 °C

Acknowledgement

Eclipse 4660 purge-and-trap courtesy of O.I. Analytical, College Station, TX.

Resolution of critical pairs, low bleed,

and high inertness make this a great

column for volatiles!

Volatile Organics Analysis

Similar to: (6%-cyanopropylphenyl)-methylpolysiloxane

For peak list, visit

www.restek.com

and enter GC_EV1169 in the search.