CALIDUS
tm
CS micro Gas Chromatograph
GC analysis for virtually any fixed gas and hydrocarbons up to C
12
for laboratory, at-line,
transportable or online use
CALIDUS Model CS -
comprised of 5 modules
A single Sample Processing Unit with a standard split/splitless injection port (1:1 up to 1:200) suitable gas and liquid
samples via either syringe through the septum injections, optional gas, liquid or headspace auto-sampler, or
automated sampling valves. The inlet includes septum purge to prevent bleed components from entering the system.
The sample is then delivered to a column switching valve for analysis on two independent column modules.
Two Programmed Temperature Column Modules (PTCM) separated by a column switching valve containing the
resistively heated steel capillary chromatography column with necessary hardware, software and electronic control to
enable temperature programming from 0.1
o
C to 5
o
C per second from 5
o
C above ambient to 180
o
C (the maximum
temperature for the valve, higher available on request) depending on the maximum temperature capability of the
column material selected. Each column module is independently controlled by the method and can be any of the
available column types.
Two independent Detector Modules incorporating either a micro Flame Ionization Detector (FID) or micro Thermal
Conductivity Detector (TCD) with the necessary hardware, software and electronic control to provide detector
temperature control, digital output signal and additionally for proper FID fuel supply pressure and auto-ignition.
The micro FID is a fully digital carbon/hydrogen bond detector using the hydrogen flame to burn the sample
components. It uses an electrometer to sense the current changes in the flame cell due to chromatographic
component elution. The data rate is 100 Hz.
The micro TCD is a fully digital, universal detector. The TCD consists of a constant temperature filament that
senses change in power required to hold the filament temperature constant when chromatographic
components elute. The power measurement is used to determine the amount of the component eluting from
the column. The data rate is 50 Hz.
CALIDUS is controlled with ChromPerfect chromatography data system fully integrated with LineUp running on a
Windows PC.
100 AEI Drive
Fairlea, WV 24901
P.O. Box 518
Ronceverte, WV 24971
Phone: 304-647-5860
Fax: 304-645-4006
E-mail:
SIGNIFICANCE AND USE
The Model CS provides a simple ultra-fast analysis (10 to 50 times faster than conventional lab or process
GC’s) of fixed gases and hydrocarbons to C
12
. Use of two different PTCMs with their own detectors
separated by the column switching valve with the appropriate column material enables leveraging the
sensitivity and selectivity differences for enhanced separations (for example heartcuts and backflushing).
Proper choice of columns enables wide boiling range and concentration ranges with a single GC. Model CS
is ideal for individual component speciation from other sample matrix components.
The analyses are used for product specifications testing, product safety, environmental measurements,
process control, catalyst protection, educational tools, spot checks of fuels and many more.
Faster, Smaller, Smarter, Easier, Greener
Upstream (E&P)
Refining
Petrochemical/Chemical
Pharmaceutical
Food & Beverage
Medical
Military
Educational
1...,8,9,10,11,12,13,14,15,16,17 19,20-21,22,23,24,25