Upstream (E&P)
Refining
Petrochemical/Chemical
Pharmaceutical
Food & Beverage
Medical
Military
Educational
CALIDUS
tm
201 micro Gas Chromatograph
GC analysis for virtually any fixed gas and hydrocarbons up to C
44
for laboratory, at-line,
transportable or online use
CALIDUS Model 201 -
comprised of 4 modules
A single Sample Processing Unit with a standard split/splitless injection port (1:1 up to 1:200) suitable gas and liquid
samples via either syringe through the septum injections, optional gas, liquid or headspace auto-sampler, or
automated sampling valves. The inlet includes septum purge to prevent bleed components from entering the
system.
Two Programmed Temperature Column Modules (PTCM) in series containing the resistively heated steel capillary
chromatography column with necessary hardware, software and electronic control to enable temperature
programming from 0.1
o
C to 5
o
C per second from 5
o
C above ambient to 350
o
C depending on the maximum
temperature capability of the column material selected. Each column module is independently controlled by the
method and can be any of the available column types.
A single Detector Module incorporating either a micro Flame Ionization Detector (FID) or micro Thermal Conductivity
Detector (TCD) with the necessary hardware, software and electronic control to provide detector temperature
control, digital output signal and additionally for proper FID fuel supply pressure and auto-ignition.
The micro FID is a fully digital carbon/hydrogen bond detector using the hydrogen flame to burn the sample
components. It uses an electrometer to sense the current changes in the flame cell due to chromatographic
component elution. The data rate is 100 Hz.
The micro TCD is a fully digital, universal detector. The TCD consists of a constant temperature filament that
senses change in power required to hold the filament temperature constant when chromatographic
components elute. The power measurement is used to determine the amount of the component eluting
from the column. The data rate is 50 Hz.
CALIDUS is controlled with ChromPerfect chromatography data system fully integrated with LineUp running on a
Windows PC.
100 AEI Drive
Fairlea, WV 24901
P.O. Box 518
Ronceverte, WV 24971
Phone: 304-647-5860
Fax: 304-645-4006
E-mail:
SIGNIFICANCE AND USE
The Model 201 provides a simple ultra-fast analysis (10 to 50 times faster than conventional lab
or process GC’s) of fixed gases and hydrocarbons to C
44
. Use of two different PTCMs in series,
for example one polar and one non-polar column material enables leveraging the selectivity
differences for enhanced separations. Secondly, using two identical PTCMs virtually doubles
the column length.
The analyses are used for product specifications testing, product safety, environmental
measurements, process control, catalyst protection, educational tools, spot checks of fuels and
many more.
Faster, Smaller, Smarter, Easier, Greener