SRI GCS Literature Detailed ex SRI website 2014 - page 819

Installation of theVacuumPump Interface on the
SRI 8610CGC
Crimp the two green/yellowwires
( from solid state relay and from
power jack ) to a single ring termi-
nal and then connect the ring ter-
minal to the chassis ground stud
with a lockwasher.
Connect the yellowwire from the
solid state relay to theA/D board
terminal labeled “D”. TheA/D
boardhas eight digital outputs (A-
H ). The yellowwire can be con-
nected to any of the digital outputs
if desired, but we normally use “D”
unless it is already occupiedwith
some other function.
Route thewires neatly through the
GC. Often therewill already be a
wire bundle and clamp through
which you can run thewires.
December 2009
Page7
1...,809,810,811,812,813,814,815,816,817,818 820,821,822,823,824,825,826,827,828,829,...870