Restek 2011/12 Chromatography Catalog - page 95

Dioxins (TCDDs) and furans (TCDFs) in fly ash on an Rxi®-5Sil MS column.
95
GC COLUMNS |
APPLICATION-SPECIFIC COLUMNS
Envi ronmental
also
available
Other Dimensions!
See
page 42
for our complete listing of
Rxi®-5Sil MS columns.
Dioxin & Furan Congeners Analysis
similar
phases
DB-5MS, VF-5ms, CP-Sil 8 Low-Bleed/MS
Rxi®-5Sil MS Columns
(fused silica)
(low polarity Crossbond® silarylene phase; selectivity close to 5% phenyl/95% dimethyl
arylene polysiloxane)
• Engineered to be a low bleed GC/MS column.
• Excellent inertness for active compounds.
• General purpose columns—ideal for GC/MS analysis of polycyclic aromatic
compounds, chlorinated hydrocarbons, phthalates, phenols, amines,
organochlorine pesticides, organophosphorus pesticides, drugs, solvent impurities,
and hydrocarbons.
• Temperature range: -60 °C to 350 °C.
The Rxi®-5Sil MS stationary phase incorporates phenyl groups in the polymer back-
bone. This improves thermal stability, reduces bleed, and makes the phase less prone to
oxidation. Rxi®-5Sil MS columns are ideal for GC/MS applications requiring high
sensitivity, including use in ion trap systems.
Si
CH
3
CH
3
CH
3
CH
3
Si
O
O
CH
3
CH
3
Si
Rxi®-5Sil MS Structure
x
y
Column
Rxi
®
-5Sil MS , 60 m, 0.18 mm ID, 0.10 µm
(cat.# 43607)
Sample
fly ash extract
Diluent:
nonane
Injection
Inj. Vol.:
1 µL splitless
Liner:
2mm Splitless liner (cat.# 20712)
Oven
Oven Temp:
120 °C (hold 1 min.) to 160 °C at 10 °C/min. to
300 °C at 2.5 °C/min.
Carrier Gas
He, constant flow
Flow Rate:
1 mL/min.
Detector
Waters AutoSpec Ultima Mass Spectrometer
Source Temp.:
280 °C
Ionization Mode: EI
Notes
Electron Ionization at 40eV
Acknowledgement Chromatogram courtesy of Karen MacPherson,
Li Shen, Terry Kolic, and Eric Reiner at the Ontario
Ministry of the Environment
GC_EV1194
1236
1237
1238
1239
2378
1278
1267
1289
TCDDs
GC_EV1195
2346
1249
1279
2347
2348
2378
2367
3467
1269
TCDFs
ID
df
temp. limits
30-Meter
60-Meter*
0.18mm 0.10µm -60 to 320/350°C
43607
0.25mm 0.25µm -60 to 330/350°C
13623
*60m, 0.18mm ID, 0.10µm column (cat.# 43607) intended for dioxin and furan analysis only.
Website :
E-mail :
TelNo : 03 9762 2034 . . . in AUSTRALIA
Mar 2011
1...,85,86,87,88,89,90,91,92,93,94 96,97,98,99,100,101,102,103,104,105,...791