GC_PH00953
1. hexane
2. 1,2-dimethoxyethane
3. trichloroethylene
4. chloroform
5. 2-hexanone
6. nitromethane
7. pyridine
8. tetralin
Residual Solvents Class 2 Mix B
Stabilwax® (G16)
Column:
Stabilwax
®
, 30m, 0.32mm ID, 0.25µm
(cat.# 10624)
Sample:
USP Stock Standard Residual Solvents
Class 2 Mix B (cat.# 36272) in 20mL headspace
vial (cat.# 24685), water diluent
Inj.:
headspace injection (split ratio 1:5), 2mm
splitless liner IP deactivated (cat.# 20712)
Inj. temp.:
140°C
Carrier gas:
helium, constant flow
Flow rate:
2.15mL/min., 35.2cm/sec.
Oven temp.:
50°C for 20 min. to 165°C @ 6°C/min.
(hold for 20 min.)
Det.:
FID @ 250°C
Headspace Conditions
Instrument:
Overbrook Scientific HT200H
Syringe temp.:
100°C
Sample temp.:
80°C
Sample equil. time.: 45 min.
Injection vol.:
1.0mL
Injection speed:
setting 8
Injection dwell:
5 sec.
Keep up with advances in melamine analysis at
Melamine Analysis Kit
............
cat. # 33254
(see page 535)
Detect melamine and related compounds by GC/MS with a complete
analysis kit. Consistent with FDA procedure and new guidance for at-risk
pharmaceutical components.
At Risk for Melamine
Meet new FDA guidance for melamine contamination
in at-risk pharmaceutical components.
Kit includes
(also sold individually)
:
• Rxi®-5Sil MS with 5 meter Integra Guard® Column
• Standards
• Derivatization Reagent
• Accessory Tubes and Syringe Filters
• Easy-to-Follow Instructions
with procedural check list
to simplify laboratory documentation!
Website :
E-mail :
TelNo : 03 9762 2034 . . . in AUSTRALIA
Mar 2011
1...,688,689,690,691,692,693,694,695,696,697 699,700,701,702,703,704,705,706,707,708,...791
Powered by FlippingBook