MRC-2013-Catalog - page 513

513
OVENS
O
600°C High Temperature
Standard features:
400
0
C, 500
0
C or 600
0
C Operating temperatures.
30, 60 & 120 Liter capacities.
301 PID controller with ramp to set point function.
Heavy duty convection fan for good uniformity.
Low thermal mass insulation for energy efficiency & rapid
heating.
Corrosion resistant, polished stainless steel interior .
2 Multi-position shelves.
Suitable for continuous operation (see options*) .
Double skin construction for cool safe outer case .
Hard wearing, zinc coated & stoved epoxy polyester
coated exterior.
LHT-Series, Laboratory High Temperature Ovens
LHT 6/60
Options:
Cable entry ports .
Over temperature protection to DIN 12-880 class 2
(recommended to protect valuable contents & for
unattended operation )*.
A range of sophisticated digital control & multi
segment programmers is available.
Optional over temperature protection
recommended for continuous operation & to
protect valuable contents.
Process timer.
RS232/RS485 communications.
Viewing window.
Chamber illumination (requires viewing window)**.
Variable speed fan.
Floor stands & stacking frames.
Routine spares kit.
Extraction fan (may alter achievable uniformity).
Stoving & curing upgrade for use with small
volumes of volatile solvent or paint fumes
(comprises over temperature protection,
extraction fan and an explosion relief panel.
**The stoving & curing option is not compatible
with the viewing door or through door
illumination options.
Model
Max.
Temp.
(
O
C)
Temp.
Stability
(
O
C)
Temp.
Uniformity
(
O
C)
@250
O
C
Heat-
up
time to
Max.
(mins)
Recovery
time to
Max.
(mins)
Dimensions
Shelves
Fitted/
Accepted
Volume
(Liters)
Max.
Power
(W)
Weight
(kg)
Power
Supply
Internal
HxWxD(mm)
External
HxWxD(mm)
LHT4/30
400 ±0.5 ±5.0
50
10 300x300x305 570x860x550 2
30 1000 73
230V
single phase
LHT4/60
400 ±0.5 ±5.0
-
16 400x400x405 670x930x670 2/3
60 1500 99
230V
single phase
LHT4/120
400 ±0.5 ±5.0
-
20 645x455x405 920x1060x650 2/4 120 2250 179
230V
single phase
LHT5/30
500 ±0.5 ±5.0
-
10 300x300x305 570x860x550 2
30 2000 73
230V
single phase
LHT5/60
500 ±0.5 ±5.0
50
16 400x400x405 670x930x670 2/3
60 2250 99
230V
single phase
LHT5/120
500 ±0.5 ±5.0
-
20 645x455x405 920x1060x650 2/4 120 3000 179
230V
single phase
or 3 phase
LHT6/30
600 ±0.5 ±5.0
70
10 300x300x305 570x860x550 2
30 2000 73
230V
single phase
LHT6/60
600 ±0.5 ±5.0
-
10(+) 400x400x405 670x930x670 2/3
60 2250 99
230V
single phase
LHT6/120
600 ±0.5 ±5.0
-
- 645x455x405 920x1060x650 2/4 120 3000 179
230V
single phase
or 3 phase
3 phase (uses 2 phases & neutral of 380/220V - 415/240V supply)
(+) Recovery to 500
O
C set -pomt