MRC-2013-Catalog - page 303

303
F
FURNACES
High Temp. Tube, 1500°C, 1600°C
STF Series, Vertical High Temperature Single Zone Tube Furnaces,
1500°C & 1600°C
STF 15/610 with over temp.
Model
STF-15/180 STF-15/450 STF-15/610 STF-16/180 STF-16/450 STF-16/610
Max. temp. (°C)
1500
1500
1500
1600
1600
1600
Heat-up time (mins)
Rate will vary with state of oxidation and age of elements
Max. o/d accessory tube (mm)
60
90
90
60
90
90
Tube/
Worktube
length(mm)
Heated
For work in air
**For modified atmosphere work
180
600
900
450
900
1200
610
1200
1500
180
600
900
450
900
1200
610
1200
1500
External H x W x D (mm)
Furnace body length (mm)
500x600x375
600
660x830x445
900
660x1130x445
1200
500x600x375
600
660x830x445
900
660x1130x445
1200
*Uniform length ±10
°
C (mm)
80
350
400
80
350
400
Max. power (W)
Supply Required (W)
-
1500
3800
5500
4200
6000
-
2500
-
6000
-
7000
T/C type
R
R
R
R
R
R
Weight (kg)
29
34
45
29
40
50
Power supply
required per phase
a1
=12A,
d1
=24A
a1
=39A,
a2
=19.5A,
c3
=25A
a1
=32A,
b1
=44A,
a2
=19.5A,
f3
=19.5A,
a3
=34A
a1
=23A
a1
=47A,
a2
=24A,
h3
=21A,
a3
=39A,
g3
=21A
a1
=50A,
a2
=25A,
h3
=25A,
i3
=27A,
d3
=43A,
e3
=46A
Continuous operating temperature is 100°C below maximum temperature. Holding power is measured at the continuous operating
temperature. Uniform length measured with end plugs fitted.
** To ensure tube end temperatures that are compatible with sealing assemblies, worktubes extending beyond the
standard length are required when working with modified atmospheres. Radiation shields may also be required.
a1
=Single 200-240V,
b1
=Single 200-208V,
d1
=Single 110-120V,
a2
=380-415V 2 phase + N,
a3
=200-240V 3 phase delta,
c3
=200-208V 3 phase+N,
d3
=200-220V 3 phase delta,
e3
=230-240V 3 phase delta,
f3
=Single 220-240V 3 phase+N,
g3
=380-415V 3 phase no N,
h3
=380-415V 3 phase + N,
i3
=400-415V 3 phase+N.
A range of tube furnaces offering the higher temperatures that are
available from silicon carbide heating elements, with the additional option
of a vertical ‘L’ stand and separate controller if preferred. The user must
select an accessory worktube specific to their application.
Requires but does not include an appropriate accessory work tube 19.5mm
to either 60mm or 90mm outer diameter
STF Series, High Temperature Single Zone Tube Furnaces, Small
1500°C & 1600°C
STF 15/610 with over temp.
Power Supplies for Silicon Carbide Furnaces
The characteristics of the control systems that
are used with silicon carbide elements result
in maximum power supply requirements that
are not as intuitively derived as those for
furnaces using other heating technologies.
For this reason a more detailed description
of the maximum power supply that is
required per phase has been included in the
specification table.
STF 15/180
Features:
1500°C & 1600°C maximum operating temperatures
Using worktubes of 60 or 90mm outer diameter
Accepts worktubes with 180, 450 or 610mm heated tube length
Silicon carbide heating elements
Horizontal configuration
Our Controller, with single ramp to set-point facility
Delayed start / process timer function as standard.
Options:
Available with ‘L’ stand kit or wall bracket for vertical or horizontal use
Available in ‘Blank-stand’ or ‘Separate-stand’ configurations
Non-permeable inner worktubes to contain modified atmosphere
Impervious inner worktubes to protect against chemical attack or
damage from thermal shock
Over-temperature protection (recommended to protect valuable
contents & for unattended operation)
Insulation plugs & radiation shields to prevent heat loss & improve
uniformity-recommended for vertical tubes
Gas injection & vacuum compatible tube end, seal assemblies
3 Phase or ‘universal’ power supply, depending upon model
8 or 20 Segment programmer
RS232 communications & graphical recorders
Vertical
1...,293,294,295,296,297,298,299,300,301,302 304,305,306,307,308,309,310,311,312,313,...738