MRC-2013-Catalog - page 299

299
F
FURNACES
Tube, 1200°C, 3 Zone
L-stand: Vertical use
TZF Series, Wire Wound Horizontal Three Zone Tube
Furnaces, 1200°C
TZF 12/75/700/3216P1
Comprising an horizontal furnace body with three
zone wire wound worktube element that is equipped
as standard with “Back to Back” zone control. It can
be used directly or with accessory worktubes.
Features:
1200°C maximum operating temperature
Excellent uniformity results from division of the heated
length into 3 equal zones each with its own controller &
thermocouple.
Provides a longer uniform zone than can be achieved in
single zone tube furnace
Power to the end zones is automatically adjusted to
compensate for heat loss
All three zones are controlled to the same set-point
Heated lengths of 400, 550, 700 or 900mm
Models accepting worktubes with outer diameters of
38mm to 90mm
Horizontally mounted on control module base.
L-stand:
Horizontal
use
Options:
Shorter end zone option to provide extended uniform
length (optionally in 2 phase configuration)
Insulation plugs & radiation shields to prevent heat loss &
improve uniformity
Gas injection & vacuum compatible worktube end seal
assemblies
Separate control module on 2 metre or 6 metre conduit
‘L’ style flexible stand option for vertical or independent
mounting
8 or 20 segment programmer
‘Retransmission of Setpoint’ zone control configuration for
programmed cooling
RS232 communications & graphical recorders
Alternative furnace sizes can be supplied upon request.
Model
TZF-12/38/400 TZF-12/65/550 TZF-12/75/700 TZF-12/100/900
Max. temp. (°C)
1200
1200
1200
1200
Heat-up time (mins)
25
45
45
120
Max. o/d accessory tube (mm)
38
65
90
90
Heated Tube length (mm)
400
550
700
900
Overall furnace length (mm)
450
600
750
950
External H x W x D (mm)
Furnace body length (mm)
430x450x375
450
525x625x360
600
525x775x360
750
525x975x360
950
*Uniform length ±5
°
C (mm)
305
390
540
754
Max. power (W)
Holding power (W)
1175
700
1817
600
2755
800
4150
1000
Thermocouple type
N
N
N
N
Weight (kg)
32
38
46
54
Power supply
Single
Single, 3 phase Single, 3 phase Single or Universal
“Universal” models are easily altered between single phase (220V), 3 phase-neutral (e.g. 380/220V) and delta
(e.g. 220V) electrical supplies.
Continuous operating temperature is 100°C below maximum temperature.
Holding power is measured at the continuous operating temperature.
Uniform length measured with end plugs fitted.
1...,289,290,291,292,293,294,295,296,297,298 300,301,302,303,304,305,306,307,308,309,...738