MRC-2013-Catalog - page 281

281
F
FUME HOOD
Cytotoxic
UV lamp
Foot switch
Inflow velocity
Work area humidity
Work area temp.
Cabinet working time
Voice Prompt
Downflow velocity
Filter Life Indicator
Fluorescent light
Waterproof Sockets
Blower
Exhaust filter pressure
Downflow filter pressure
Filter Working Time
UV Working Time
Time Reservation
Model
BSC-CY-4
External size (WDH mm)
1220x600x630 mm
Internal size (WDH mm)
1370x760x2100 mm
Tested Opening
Safety Height < 200mm
Max Opening
480mm
Average Airflow Velocity
Inflow: 0.46 m/s
Downflow: 0.33 m/s
Internal Work Area, Space
0.73m
2
Exhaust Volume With
Thimble Duct
CBV Exhaust Volume 611 m
3
/h (360 cfm)
Static Pressure at CBV 39 Pa / 0.15 in H20
Air Supply Filter
ULPA filter, efficiency 99.999% between 0.1~0.2 um
Exhaust Filter
2 pieces ULPA filter, efficiency 99.999% between
0.1~0.2 um
Front Window
Two-layer laminated toughened glass
Noise
62 dB
UV Lamp
30W*1
fluorescent Lamp
28W*2
Illumination
1000 Lux
Consumption
700 W
Water & Gas Tap
Water Tap* 1; Gas Tap*1
Socket
2 Waterproof Socket: 2x250W (Max)
Display
LCD Display
Control System
Microprocessor
Material
Work Area: 304 stainless steel
Main Body: Cold-rolled steel with anti-bacteria
powder
Caster
Universal wheel
Power Supply
110/220V±10%, 50/60Hz
Standard Accessory
Remote control, Foot switch, UV lamp,
Fluorescent lamp, Waterproof socket
Optional Accessory
Armrest, Air curtain tester, Formanlin
fumigation sterilizer, Infrared Sterilizer
Gross Weight
400kg
Package Size (WxDxH)
1570x950x2220 mm
BSC-CY-4
The unique demands of handling and preparing cytotoxic drugs for use in
chemotherapy require a specialized cabinet. As cytotoxic drugs cannot be
inactivated by chemical decontamination, Class II biosafety cabinets should
not be used. With this in mind MRC has developed a highly specialized
range of cabinets designed especially for handling these potentially
dangerous drugs
The MRC Cytotoxic Safety Cabinet is the premium solution for cytotoxic/
antineoplastic drug processing, providing the highest level of patient,
pharmacist and environmental protection. This revolutionary product builds
on MRC’s experience of more than 10 years as an expert in biosafety
containment technology.
BSC-CY-4, Class
II
Biosafety Cabinet
A
Room Air
Polluted Air
HEPA Filter Air
Protected Area
A
B
Unprotected Area
1...,271,272,273,274,275,276,277,278,279,280 282,283,284,285,286,287,288,289,290,291,...738