MRC-2013-Catalog - page 280

280
FUME HOOD
F
Biological
BSC11
II
B2X/BSC13
II
B2X/BSC15
II
B2X, Class
II
Model
BSC11
II
B2X BSC13
II
B2X BSC15
II
B2X BSC18
II
B2X
External size (WDH mm)
1100x850x2200
1300x850x2200
1500x850x2200
1800x850x2200
Internal size (WDH mm)
900x600x660
1100x600x660
1300x600x660
1600x600x660
Tested Opening / Max Opening
Safety Height
200mm / 400mm
low velocity / Down Flow velocity
0.53m/s / 0.33m/s
Pre-filter / Filter
Washable / Two Hepa filter with 99.999% efficiency at 0.3 micros
Front Window
Two-layer toughened glass >5mm
Noise level
65 dB
Material
Work zone interior is made of 304 stainless steel.
Body: expoxy coated cold rolled steel
Base Height
635mm
Control system / Caster Wheel
Microprocessor / Directional Wheel
Display
VFD display
Airflow System
0% air recirculation, 100% air exhaust
Exhaust Duct
Material PVC Diameter: 300mm
Clean level / Lightning
100 / 800Lux
UV lamp
20W x 1
20W x 1
30W x 1
30W x 1
Fluorescent lamp
21W x 1
21W x 1
28W x 1
30W x 1
Water & Gas Tap / Waterproof Socket
Water Tap: 1, Gas Tap: 1 / 2
Electrical / Power (W)
110V~ 240V/50Hz~60Hz (optional) / 1500
Net weight (kg)
275
320
350
375
Gross weight (kg)
350
360
440
470
Package (mm)
Main Body
1250x1050x1870 1450x1050x1870 1640x1040x1860 1940x1050x1860
Exhaust Blower
1100x1000x900
700x600x1000
700x600x1000
1060x970x580
Standard Configuration
Body, Remote control, Foot switch, UV lamp & Fluorescent Lamp,
4 meters, Exhaust duct, 1 exhaust blower, Base Stand
Optional Accessory
Armrest, Air curtain tester, Formalin fumigation sterilizer, Ammonium
hydrogen carbonate neutralizer
Exhaust Blower
Remote Control
LCD Display
VFD Display
Class
II
B2 Biosafety Cabinets feature downflow air drawn from within
the laboratory. No downflow air is drawn from the cabinet exhaust air. All
downflow & inflow air is exhausted through a HEPA filter without recirculation
within the cabinet. All contaminated ducts and plenums are maintained
at negative pressure. The Type B2 Biosafety Cabinet may be used for work
with volatile toxic chemicals and radionuclides as required as adjuncts to
microbiological studies.
BSC11
II
B2X
Room Air
Polluted Air
HEPA A Filted Air
Protected Area
A
B
Unprotected Area
1...,270,271,272,273,274,275,276,277,278,279 281,282,283,284,285,286,287,288,289,290,...738