MRC-2013-Catalog - page 26

26
MGB Series of microbalances have been
designed to meet the high requirements of
mass measurements with the highest
precision. Measurement reliability and
accuracy is ensured by internal calibration.
Microbalances consist of two major parts
(an electronic system and a precise
mechanical measurement system in a
separate enclosure). This solution eliminates
the temperature influence and separates
from shocks and vibrations caused by users
operating software.
All the elements of the balance are made
of glass and steel which eliminates the
influence of electrostatics on the weighing
process.
Filling
Chckweighing
Percentage
%
Statistics
Infrared Sensors
GLP Procedures
LP
Electronic level
indicator
ALARM function
Graphic level
indicator
Programmable
acceptable tilts
Infrared proximity
sensors
PRINT function
TARE function
Opening weighing
chambers
Sensors’ sensitivity
adjustment
Optional accessories:
Anti-vibration table
Printer
Set for the determination of air
Foot tare and print buttons
Additional attachment for Pipettes
calibration
Professional weighing table
Ultrasonic air ionizer with humidifier
Standard mass
Computer software
Air Buoyancy Correction
HI
LOW
OK
Micro
BALANCES
MGB
Model
MGB-2
MGB-3
MGB-5
MGB-11
MGB-21
Max capacity
2g
0.8/3g
5g
11g
21g
Readability
1 μg
1/10 μg
1 μg
1 μg
1 μg
Tare range
-2g
-3g
-5g
-11g
-21g
Repeatability*
1 μg
1 μg
1.6 μg
2.0 μg (to 5g)
2.5 μg (5g-11g)
2.0 μg(to 5g)
2.5 μg(5g-11g)
3.0 μg (11g-21g)
Linearity
±3 μg
±3 μg
±5 μg
±6 μg
±7 μg
Eccentric load deviation
3 μg
3 μg
5 μg
6 μg
7 μg
Sensitivity offset
1.5 x 10
-6
x Rt
3 x 10
-6
x Rt
4 x 10
-6
x Rt
Sensitivity temperature drift
1 x 10
-6
/ °C x Rt
Sensitivity stability
1 x 10
-6
/ Rok x Rt
Minimum weight (USP)
3mg
3mg
6.3mg
6.3mg
6.3mg
Minimum weight (U = 1%, k = 2)
0.2mg
0.2mg
0.4mg
0.4mg
0.4mg
Pan size
Ø
16mm
Ø
16mm+
Ø
60mm
(weighing pan for filters)
Ø
26mm
Weighing chamber dimensions
Ø
90 x 90mm
Stabilization time
5 s
Adjustment/Calibration
automatic (internal)
Working temperature
+18 ° - +30 °C
Change rate of working temp.
±0.3 °C/h (±1 °C/8h)
Atmospheric humidity
45% - 65%
Change rate of atmospheric humidity
±1%/h (±4%/8h)
Interface
2xUSB, 2xRS 232, Ethernet, 2in/2out (digital)
Power supply
230V / 11V AC
Display
5.7’’ touch screen
MGB-Series, Micro Balances
*Repeatability as a standard deviation from 10 weighing cycles
B