D
COD Thermoreactor
Model
DBR-001 COD REACTOR
Read out
4 Digital LED of set temperature, set time, process temperature, residual time display
Timer
1~999 minutes with audible alarm and automatic shut off or continuous operation
Accuracy
±2°C (at 150°C)
Temperature
Fixed 150°C (COD program) 60°C~200°C, adjustable
Safety
Hot top indicator (flashing when block temperature is over 70°C
(Automatic shut off when block temperature is over 210°C)
Block capacity
25 of 16mm
Ø
sample vials, Option block(with 8x20mm
Ø
and 6x25mm
Ø
vial holes) is available
Structure
Housing: stainless steel with power paint coating, Block: Anodized aluminum
Power
AC110 or AC220V, 50/60Hz, 220Watt
Dimensions
W18.9xH11.6xD31.5mm
Weight
Net 4.5kg Gross 6.1kg
DBR-001
25x16mm
Ø
block Included
Features:
Approved by ERA (method 410.4)
of USA and most of the countries
in the world.
Safer and easier than traditional
open reflex method.
Save lab top space, power and
water consumption.
Reduce waste & maximize safety.
Unique COD program makes it
truly automatic.
For The Digestion Of:
COD (150°C)
TOC (120°C)
Total chromium (100°C)
Total nitrogen (100°C)
Total phosphate (100°C)
Chemical digestion of the samples is a prerequisite such as: tests,
determination of COD, TOC, total phosphate, nitrogen etc.
The required temperature & reaction time are programmed using the membrane
keypad on the front of the reactor. When digestion is complete, the reactor
automatically switches off & sounds an alarm. DBR-001 COD reactor is designed
to meet USEPA 410.4 method. It can be used to digest closed I micro reflux COD
vial and other samples which diameter is 16mm.
Further, it is safer and easier than traditional open macro reflex method.
Unique COD program let DBR-001 to be a truly automatic reactor.
When the reactor is powered on, just click the start/stop button, then it starts to
heat to 150°C and keep for 2 hours then stop heating with an audio alarm.
Samples are ready to be measured by colorimeter when they cool down.
DBR-001 COD reactor is compatible with any COD detectors which use 16mm
COD vials. In additional to COD program, DBR-001 is a microprocessor controlled
block heater, with temperature range from 60°C to 200°C and timer from 0 to
999 minutes or continuous operation, and 3 extra self-made heating programs
can be stored.