CARRIER

GAS

GENERATOR

HIGH INTENSITY

HIGH FREQUENCY

CELL OPTICAL

FILTER

PHOTO-DIODE AMPLIFIER

SPECIFICATIONS:

CARRIER GAS

Argon and Helium

POWER

80 to 240 VAC, 50-60 Hz

GAS CONNECTIONS

1/16” (can be customized)

OPERATION OUTLET PRESSURE

Atmospheric or Vacuum

OPERATING TEMPERATURE

10°C to 50°C (in stable environment)

FILTER

10u SS particle filter on the gas inlet

DETECTOR SIGNAL OUTPUT

BNC Coaxial type (can be customized)

CONNECTION

POWER CONSUMPTION

Max 10 Watts

OUTPUT VOLTAGE

0-5 Volts (can be customized)

PRINCIPLE OF PLASMA EMISSION DETECTOR (PED)

The PED is a quartz cell with a

unique design submitted to a

high intensity and frequency

electromagnetic field.

The principle based on spec-

troscopic emission cell is not a

new technique, but the

characteristics of the Plasma-

Detek system that make it

stable and efficient are the

frequency, the intensity as well

as the mechanical and electro-

des design.

A luminous phenomenon,

called electroluminescence, is

created and is used as emis-

sion technique to quantify

analytes.

When the carrier gas is

ionized, spectral lines are

emitted and detected by an

optical system including filter

and photo-diode. The emission

varies for each substance that

is brought along with the

carrier gas.