Chapter: INSTALLATION
Topic: PowerSupplies and SpaceRequirements
Once the equipment has been removed
from all the packing material, check the
contentsof thecontainer against thepacking
slip and make sure everything listed is
included. If any item(s) have been omitted
oraremissing, contact SRI Instruments for
locationand orreplacement
ofthe
item(s).
The SRI
model
8610C gas
chromatograph requires ACpower ateither
110 VAC at 60 Hertz or 220 VAC at 50
Hertz, dependingon the ACpower ordered.
Both AC power supplies are equippedwith
a three prong grounded outlet (see diagrams
to the
right).
Proper grounding is required
for safe operation.
Do not disable the
ground prong under any circumstance.
These plug configurations are for EIA
standard U.S. outlets. It maybe necessary
to replace the plug provided with a local
standard plug.
o •
llOVAC
220VAC
A standard SRI 8610C GC measures
18.5" X 14.5" X 12.5" and requires a
minimum counter space of
28" X 22" X
23.5" for proper operation (see diagram to
the right). Roughly 8" of clearance beside
the left side control panel is needed fordata
cable, gas line and powerswitchaccess. 6"
of clearance to the rear of the GC and 11"
of clearance above theGCisrequired. This
will provide adequate access to the column
oven formaintenance and provide space for
proper GC ventallation, To the front and
right side, 1.5" of clearance should be
adequate to prevent the GC from coming
into contact with surrounding objects or
falling off the counter. The right side of the
GC does contain general information on
your instrument and some operators may
want additional clearance foreasy reference.
The front control paneloftheGC should be
easily accessible in order to properly
monitor digital display and control operating
conditions, as well as providing access to
theinjectionportforsample injections.