SRITN12 - page 414

ExpectedPerformance
Thefollowing threechromatogramsarefrom theFID inaSRIGCwithaPTV injectorupgrade.The liner
waspackedwith0.1gramsofTenax-GR
TM
adsorbent. All three25minuterunsutilized thesame temperature
andeventprograms. In thefirstone,a1µL2000ppmC
10
-C
28
samplewas injected through thePTV injector.
Inthesecondchromatogram,thesamesamplewasdiluted1:100,then100µlinjected,achievingresultsconsistent
with the first run,anddemonstrating the
highvolume liquid injectioncapabilityof
the PTV injector. In the third
chromatogram,100µLofmethanolwas
injectedasablank,resultinginasmallhump
between the 4 and 7minutemarks and
miniscule peaks which correspond to
contaminants in themethanolblankand
bleedfrom theTenax-GR
TM
.
PTV=110
o
C (3min) to275
o
C
Temperatureprogram:
Initial
Hold Ramp Final
110
o
C 7.00 15.00 270
o
C
270
o
C 20.00 0.00 270
o
C
Events: (A=split vent)
Time Event
0.00 ZERO
3.00 AOFF
3.10 BON
3.20 CON
5.00 BOFF
Temperature programs & events for all 3 runs:
1
2
3
Chromatogram 1 Results:
Component
Retention Area
Solvent
0.866 84953.1370
C10
5.366
5299.9150
C12
7.300
5034.0980
C14
10.233
4814.2000
C16
12.450
4600.0300
C18
14.216
4436.1780
C20
15.750
4528.2890
C22
17.150
4570.0975
C24
18.483
4778.9380
C26
20.033
4863.4290
C28
22.216
4135.4760
Total 132013.7875
Chromatogram 2 Results:
Component
Retention Area
Solvent
0.450 499472.8740
C10
5.433
2258.5340
C12
7.366
2614.0540
C14
10.266
3813.8985
C16
12.483
3924.8340
C18
14.266
3939.9080
C20
15.800
3933.0400
C22
17.200
4660.5860
C24
18.516
4737.3130
C26
20.083
4174.2920
C28
22.266
3260.1120
Total 536789.4455