SRITN12 - page 255

GC INJECTORS
HTA Autosampler
2. Remove the bottom cover from theGC by unscrewing and removing the six phillip’s head screws
that secure it to theGC. Rock theGCgentlyonto it’s back andpull the bottom cover off towardyou,
as shown. While theGC is in thisposition, remove thegreenandyellowgroundwire from thebottom
cover.
TheGC groundwire is on the left-hand side of
the bottom cover
1. TheHTAand theGCwill arrive in twoseparatepackages: theGC inonecrate, theHTAautosampler
andbase (includinganewGCbottomcover) inanother. Unpack them, but leave theblack strapon the
HTA(as shownbelow) until it is set upand ready tooperate. Lay theHTAon it’s sideas shown. Place
theGC on a level surface so you canwork on it.
3. Place thenewbottomcoverandHTA
stand base next to the GC, and attach
the green and yellowwire to the bolt
with the washer and nut from the old
GC bottom cover.
HTAautosampler
GC bottom cover and
HTAstand base