SRITN12 - page 261

Chapter: INSTALLATION
Topic: PowerSupplies and SpaceRequirements
Once the equipment has been removed
from all the packing material, check the
contentsof thecontainer against the packing
slip and make
sure
everything listed is
included. If any item(s) have been omitted
orare missing, contact SRI Instruments for
locationand orreplacement
ofthe
item(s).
The SRI
model
8610C gas
chromatograph requires ACpower ateither
110 VAC at 60 Hertz or 220 VAC at 50
Hertz, dependingon the AC power ordered.
Both AC power supplies are equippedwith
a three pronggrounded outlet (see diagrams
to the right). Proper grounding is required
for safe operation.
Do not disable the
ground prong under any circumstance.
These plug configurations are for EIA
standard U.S. outlets. It maybe necessary
to replace the plug provided with a local
standard plug.
o •
110
VAC
220VAC
A standard SRI 8610C GC measures
18.5" X 14.5" X 12.5" and requires a
minimum counter space of
28" X 22" X
23.5" for proper operation (see diagram to
the right). Roughly 8" of clearance beside
the leftside control panel isneeded for data
cable. gas line and powerswitch
access.
6"
of clearance to the rear of the GC and 11"
of clearance above theGCisrequired. This
will provide adequate access to the column
oven formaintenance and provide space for
proper GC ventallation. To the front and
right side, 1.5" of clearance should be
adequate to prevent the GC from coming
into contact with surrounding objects or
falling off the counter. The right side of the
GC does contain general information on
your instrument and some operators may
want additional clearance foreasy reference.
The front control paneloftheGC should be
easily accessible in order to properly
monitor digital display and control operating
conditions, as well as providing access to
the injectionportforsample injections.