High ResolutionGC/MS Separations of
Dioxin and Furan Congeners
Using Restek's NewRtx
®
-Dioxin2Capillary GCColumn
the
2003
vol.4
Advantage
Innovators of HighResolution
Chromatography Products
in this issue
Dioxin and Furan
Congeners GC/MSAnalysis
Pg. 1–2
Chlorophenoxyacid
Herbicides HPLCAnalysis
Pg. 3,5
Extracolumn Volume and its Effects in
Gradient HPLC
Pg. 4–5
Replacement Parts for HPLCSystems
Pg. 6–7
Faster GCAnalysis of Purgeable Organics
Pg. 8–9
NewReferenceMaterials for Forensic and
Environmental Analyses
Pg. 10
Bulk HPLCSilica and Bonded
Phase Packings
Pg. 11
Peak Performers—Avoid Septum
Problems
Pg. 12–13
Cool Tools—Super-Clean
Gas
Filters
Pg. 14–15
By FrankDorman, Ph.D., Director of Technical Development
Resolves 2,3,7,8-substituted congeners from each other and from non-toxic congeners.
Resolves furan congeners from chlorodiphenyl ethers.
Stable to 320ºC, for longer column life.
An accurate GC analysis of dioxin and furan con-
geners is a challenge. Separation of the toxic con-
geners (configurations with substitutions at the 2, 3,
7, and 8 positions) from the non-toxic congeners is
difficult on
almost any sta-
tionary phase.
Most laborato-
ries perform an
initial analysis
using a 5%
diphenyl / 95%
dimethyl poly-
siloxane col-
umn (e.g.,
Rtx
®
-5) to
obtain reasonable estimates of concentrations for the
2,3,7,8-substituted congeners. For some of the target
congeners, this quantitation is biased toward high
values, due to coelutionwith non-toxic congeners.
Asmany as five non-toxic TCDFs can coelutewith
2,3,7,8-tetrachlorodibenzofuran, for example, in an
analysis on a 5% diphenyl / 95% dimethyl polysilox-
ane (5-type) column.
Because of this coelution issue, many laboratories
use confirmation columns, most commonly high
cyanopropyl (225-type) stationary phases (e.g.,
Rtx
®
-225, Rtx
®
-2330), in order tomore accurately
quantify the toxic congeners. Unfortunately,
cyanopropyl columns exhibit poor thermal stability,
and therefore column lifetimes are short.
Sincemost methods for analysis of dioxins and
furans include extensive sample extract cleanup, and
high-resolutionmass spectrometry, a primary
requirement of the ideal analytical column is com-
plete separation of the toxic dioxin and furan con-
geners from each other. Additionally, it is desirable
for the column to have high thermal stability and
long lifetime.
With these characteristics inmind, Restek chemists
developedRtx
®
-Dioxin2 capillary GC columns. These
new columns completely resolve the 2,3,7,8- substi-
tuted congeners from each other, and from the non-
toxic congeners as well. Figure 1 shows the separa-
tion of the tetrachlorodibenzodioxins on a 60m x
0.25mm ID x 0.25µmRtx
®
-Dioxin2 column. 2,3,7,8-
TCDD is well resolved from the other congeners in
this group and can be quantified accurately. The col-
umn also is available in an alternative format com-
monly used for this analysis: 40m x 0.18mm ID x
0.18µm. Either column is stable to 320ºC.
We’ve
expanded
ourwebsite!
•New Features.
• Fast searches.
• Easy navigation.
Wewelcome
yourvisit.
Figure1
—2,3,7,8-Tetrachlorodibenzodioxin canbe resolved fromotherTCDD congenersbyusing
anRtx®-Dioxin2 column.
60m, 0.25mm ID, 0.25µmRtx
®
-Dioxin2
(cat.# 10758)
Oven temp.:
130°C (hold 1min.) to
205°C@ 45°C/min. to 305°C@
6°C/min. (hold 30min.);
Dead time:
2.89min.;
Carrier gas:
helium at
1.5mL/min., constant flow
Chromatogram courtesy of KarenMacPherson and Eric Reiner, OntarioMinistry of the Environment, Etobicoke, Ontario, Canada.
Native TCDDs
321.8936m/z
1,2,3,4-TCDD-C13
2,3,7,8-TCDD-C13
C13 labels
331.9386m/z
GC_EV00702
Restek
Exclusive
!
new
2,3,7,8-TCDD