restekadv07 - page 154

with data pack
with data pack
• 13 •
Table I
Small relative standard deviations for calibration factors show a DrilledUniliner
®
inlet liner efficiently
transfers the sample to the column.
Relative Standard Deviation (%)
CLPesticides
CLPesticides2
Column
Column
α
-BHC
7.0
7.5
γ
-BHC
3.3
3.8
β
-BHC
10.2
9.6
δ
-BHC
6.0
7.1
heptachlor
4.2
10.9
aldrin
2.3
1.2
heptachlor epoxide
10.7
8.6
γ
-chlordane
6.8
6.7
α
-chlordane
8.3
6.7
4,4' DDE
2.3
3.3
endosulfan I
9.2
8.3
dieldrin
7.6
6.6
endrin
4.9
5.3
4,4' DDD
2.7
4.1
endosulfan II
9.9
9.7
4,4' DDT
3.8
2.4
endrin aldehyde
12.3
13.3
methoxychlor
10.2
10.8
endosulfan sulfate
9.3
10.6
endrin ketone
7.9
5.3
DI Liners for Agilent 5890/6890 GCs
(For 0.25/0.32/0.53mm IDColumns)
ID*/OD&
Length (mm)
cat.#/price
ea.
cat.#/price
5-pk.
Drilled Uniliner
®
4.0 ID
6.3OD x 78.5
21054
21055
Siltek
Drilled Uniliner
®
4.0 ID
6.3OD x 78.5
21054-214.1 21055-214.5
Siltek
1mmDrilled Uniliner
®
1.0 ID
6.3OD x 78.5
21390-214.1 21391-214.5
Standard:
cat#. 32292, 8/16/80µg/mL
in hexane:toluene (1:1)
Calibration curve standards:
5/10/50ng/mL in hexane
20/40/200ng/mL in hexane
80/160/800ng/mL in hexane
Uniliner
®
Inlet Liners for Agilent GCs
Chlorinated pesticides listed in US EPAMethod 8081.
Stx
-CLPesticides
(FusedSilicawithSiltek
deactivation)
ID
df (µm)
temp. limits
15-Meter
30-Meter
0.25mm
0.25
-60 to 310/330°C
11540
11543
0.32mm
0.50
-60 to 310/330°C
11541
11544
0.53mm
0.50
-60 to 310/330°C
11542
11545
Stx
-CLPesticides2
(FusedSilicawithSiltek
deactivation)
ID
df (µm)
temp. limits
15-Meter
30-Meter
0.25mm
0.20
-60 to 310/330°C
11440
11443
0.32mm
0.25
-60 to 310/330°C
11441
11444
0.53mm
0.42
-60 to 310/330°C
11442
11445
*Nominal ID at syringe needle expulsion point.
decachlorobiphenyl
2,4,5,6-tetrachloro-
m
-xylene
200µg/mL each in acetone, 1mL/ampul
Pesticide SurrogateMix
Each
5-pk.
10-pk.
32000
32000-510
32000-500
32000-520
32100
aldrin
8µg/mL
α
-BHC
8
β
-BHC
8
δ
-BHC
8
γ
-BHC (lindane)
8
α
-chlordane
8
γ
-chlordane
8
4,4'-DDD
16
4,4'-DDE
16
4,4'-DDT
16
dieldrin
16
endosulfan I
8
endosulfan II
16
endosulfan sulfate
16
endrin
16
endrin aldehyde
16
endrin ketone
16
heptachlor
8
heptachlor epoxide (B)
8
methoxychlor
80
In hexane:toluene (1:1), 1mL/ampul
Organochlorine
PesticideMix AB #2
Each
5-pk.
10-pk.
32292
32292-510
32292-500
32292-520
32392
The Drilled Uniliner
®
inlet liner is the
first inlet liner to allow direct injections
in EPC systems!
INNOVATION
linearity values (%RSD) obtained using a Drilled
Uniliner
®
liner and Rtx
®
-CLPesticides columns.
Equivalent results are obtainedwith
Stx
-CLPesticides columns.
When using a Drilled Uniliner
®
liner, the efficient
transfer of sample from the injection port to the
columnmight allowmore nonvolatilematerial to
enter the column. Formany samples this will not be
an issue, especially if solid phase cleanup is per-
formed. Using a guard columnwill helpmaintain
the analytical column. When necessary, the guard
column can be trimmed to remove the portion
where the nonvolatilematerial collects (usually the
first 6-12 inches).
Conclusion
TheDrilledUniliner
®
liner provides the advantages
of both direct and splitless injection liners. The liner
provides amore inert sample pathway to transfer
the sample from the injector to the analytical col-
umn, and it helps eliminate injection port discrimi-
nation. ADrilledUniliner
®
liner will reduce the
detection limits for injectionsmade on a splitless
injection port, without the need for an on-column
injector.
On l i ne
at
Ordering