restekadv07 - page 153

Figure 2
Stx
-CLPesticides andStx
-CLPesticides2 columns provide rapid, excellent separation of chlorinated pesticides, and a Siltek
-deactivatedDrilled
Uniliner
®
inlet liner helps ensure high responses for sensitive analytes.
• 12 •
by Gary Stidsen, InnovationsManager
Inert Inlet System Improves
Responses for Chlorinated
Pesticides
Using theDrilledUniliner
®
Inlet Liner
Inert sample path eliminates injection port discrimination.
Reduce detection limits by using a splitless injection port without an on-column
injector.
For years, chemists analyzing chlorinatedpesticides
have triedmany different injection techniques in
attempts to find the best balance between inertness
and ability to contendwith sample contamination.
Cool on-column, split, splitless, and direct injec-
tion, and variations of these injection techniques,
are used today.
Now, a speciallymodified injectionport liner, devel-
opedby Restek chemists, reduces sample contact
with activemetal parts in split/splitless injectionports.
ThisDrilledUniliner
®
liner, shown in Figure 1, gives
the benefits of bothdirect injection and splitless
injection. The advantage of this liner is that the col-
umn is connected to the liner by a press-fit connec-
tion, thus preventing the sample from contacting the
metal at the bottomof the injectionport. Also, the
hole on the side of the liner allows the purge flow to
escape from the linerwhen the injectionmode is
switched from splitless to split.
Inertness
With the Drilled Uniliner
®
liner, the sample is trans-
ferred directly from the injection port to the col-
umn and contacts only this glass inlet liner. The
configuration allows the sample to be "funneled"
into the column entrance, thereby eliminating the
need for vaporization aids such as fused silica
wool. The test probes endrin and 4,4'-DDT are
good indicators of injection port inertness. Analyses
of the performance evaluationmixture (PEM) show
low breakdown of these compounds when using
this liner.
Formaximum inertness, we recommend using the
Drilled Uniliner
®
liner in combinationwith an Stx
-
CLPesticides capillary column, as shown in Figure
2. Siltek
surface deactivation in these columns fur-
ther ensuresmaximum responses for the labile pes-
ticides endrin, 4,4'-DDT, andmethoxychlor. A spe-
cial polymer formulation designed for organochlo-
rine pesticides enables you to achieve excellent sep-
aration of the 22 chlorinated pesticides in Figure 2
in less than 15minutes. Comparable analyses on
Rtx
®
-CLPesticides columns take up to 24minutes
(Figure 2, inset).
Calibration
Linearity and continuing calibration checks for the
chlorinated pesticides also are critical parameters
that must bemonitored. Table 1 indicates typical
1. 2,4,5,6 tetrachloro-
m
-xylene (IS)
2.
α
-BHC
3.
γ
-BHC
4.
β
-BHC
5.
δ
-BHC
6. heptachlor
7. aldrin
8. heptachlor epoxide
9.
γ
-chlordane
10.
α
-chlordane
11. 4,4'-DDE
12. endosulfan I
13. dieldrin
14. endrin
15. 4,4'-DDD
16. endosulfan II
17. 4,4'-DDT
18. endrin aldehyde
19. methoxychlor
20. endosulfan sulfate
21. endrin ketone
22. decachlorobiphenyl (IS)
Oven temp.:
110°C (hold 1min.) to 245°C@20°C/min.
to 300°C@6°C/min.
Inj. & det. temp.:
210°C / 310°C
Carrier gas:
helium
Dead time:
0.8min. @120°C
Inlet liner:
Siltek
DrilledUniliner
®
liner (cat.# 21055-214.5)
Inj.:
1µL direct injection of 20/40/200ng/mL std.
concentration in hexane
Make-up gas:
nitrogen
1
2 3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
30m, 0.32mm ID, 0.5µm
Stx
-CLPesticides
(cat.# 11544)
30m, 0.32mm ID, 0.25µm
Stx
-CLPesticides2
(cat.# 11444)
1
2
3
4
5
6
7
8
9
10
12
11
13
14
15
16
1718
19
20
21
22
GC_EV00512
Figure 1
The drilled hole in a Uniliner
®
injection port
linermakes direct injection possiblewith
EPC systems by equalizing pressure in the
injection port.
8
10
12
14min.
8
10
12
14min.
6
9
10
11
12
13
14
15
16
17
18
19
20
1
2
3
4
5
6
7
8
21
22
min. 10
24
GC_EV00075
On-column concentration:
16–160pg organochloride
pesticidemix AB#2
(cat.# 32292)
Oven temp.:
120°C (hold 1min.) to 300°C
(hold 10min.)@9°C/min.
Inj. port:
Direct, Uniliner
®
liner
(cat.# 20335), at 200°C
Detector:
ECD, 300°Cwith Anode Purge
Dead time:
1.9min.
Head pressure:
8.7psi (constant)
Flow rate:
1.3mL/min. @120°C, helium.
30m, 0.32mm ID, 0.25µm
Rtx
®
-CLPesticides2
(cat.# 11324)
Traditional analyses on
Rtx
®
-CLPesticides columns take
up to 24minutes.