Table of Contents Table of Contents
Previous Page  180 / 413 Next Page
Information
Show Menu
Previous Page 180 / 413 Next Page
Page Background

• 6 •

2008 vol. 2

Acetamide herbicides are used in large quantities to suppress weed growth in corn and soybean fields. However, due to the polar

nature of ethanesulfonic acid (ESA) and oxanilic acid (OA) degradation products, contamination of drinking water sources is a con-

cern. EPA Method 535 is designed to monitor drinking water for ESA and OA breakdown products of these herbicides.

Chromatographic analysis is extremely important for this method because two analytes, alachlor ESA and acetochlor ESA, are

isomers that share the same mass spectral multiple reaction monitoring (MRM) transitions, and thus must be separated chromato-

graphically.

Resolution of all Method 535 analytes, including alachlor ESA and acetochlor ESA isomers, can easily be accomplished using Restek's

full line of Method 535 products, which includes reference standards, solid phase extraction (SPE) cartridges, and HPLC columns

that meet method guidelines. In the procedure shown here, 6mL CarboPrep™ 90 SPE cartridges were used for sample preparation,

both to help extend the lifetime of the analytical column as well as to prevent matrix enhancement or suppression. LC/MS/MS analy-

sis was performed on an Ultra C18 column coupled to an Applied Biosystems API 3200™ LC/MS/MS system equipped with a

TurboIonSpray® probe for the Turbo V™ source.

One Stop Shop for EPA Method 535

Reliably Analyze Acetamide Herbicide Degradates by LC/MS/MS

By Jason Thomas, Innovations Chemist, Julie Kowalski, Ph.D., Innovations Chemist, and Christopher Borton, Applied Biosystems

• Full package: reference standards, SPE cartridges, and HPLC columns.

• Chromatographic resolution of alachlor ESA and acetochlor ESA isomers.

• Meet method requirements, with superior sensitivity.

Figure 1

Easily resolve Method 535 herbicide degradates on an Ultra C18 HPLC column.

alachlor OA

acetachlor OA

1. alachlor ESA

3. acetachlor ESA

4. metalochlor ESA

2. metalochlor OA

dimethachlor ESA (SS)

butachlor ESA

Sample:

acetamide herbicides

Inj.:

25µL

Conc.:

50ng/mL

Sample diluent: 5mM ammonium acetate in water

Column:

Ultra C18

Cat.#:

9174312

Dimensions:

100mm x 2.1mm

Particle size:

3µm

Pore size:

100Å

Conditions:

Mobile phase:

A: 5mM ammonium acetate

B: methanol

Time:

%B

0.0

20%

4.0

30%

10.0

30%

15.0

50%

17.0

85%

18.0

85%

18.1

20%

28.0

20%

Flow:

250µL/min.

Temp.:

ambient

Excellent response at

low concentrations!

Fully resolve alachlor ESA

and acetochlor ESA isomers.

Det.:

API 3200™ LC/MS/MS system

Ion Source:

TurboIonSpray

®

probe for Turbo V™ source (negative)

Mode:

MRM

Curtain gas:

25

Collision gas:

medium/5

IonSpray voltage:

-4,500

Temperature:

500°C

Ion source gas 1:

50

Ion source gas 2:

50

Interface heater:

on

Dwell time:

50

Vertical probe position: 2

Analytes

RT (min) Quantifer MRM Qualifier MRM

dimethachlor ESA 7.9

300/80

300/121

alachlor OA

14.5

264/160

264/158

acetochlor OA 14.8

264/146

264/144

1. alachlor ESA

16.1

314/80

314/121

2. metolachlor OA

16.4

278/206

278/174

3. acetochlor ESA 16.4

314/80

314/121

4. metolachlor ESA 16.9

328/80

328/121

butachlor ESA

19.7

356/80

356/121

LC_EV0477

Environmental