Column: Rxi

®

-624Sil MS, 30 m, 0.32 mm ID, 1.80 µm

(cat.# 13870)

Sample: Residual Solvents Class 2 - Mix A (cat.# 36271)

Complete analytical conditions for chromatogram

GC_PH1161 available at

www.restek.com

GC_PH1161

Figure 1

Rxi®-624Sil MS columns reliably resolve Class 2 Mix A compounds.

Peaks

RT (min.)

Conc.

(µg/mL)

1. Methanol

2.281 25.00

2. Acetonitrile

4.009

3.42

3. Dichloromethane

4.313

5.00

4.

trans

-1,2-Dichloroethene 4.798

7.83

5.

cis

-1,2-Dichloroethene 7.028

7.83

6. Tetrahydrofuran

7.706

5.75

7. Cyclohexane

8.708 32.33

8. Methylcyclohexane

14.099 9.83

9. 1,4-Dioxane

15.054 3.17

10. Toluene

22.018 7.42

11. Chlorobenzene

26.570 3.00

12. Ethylbenzene

26.837 3.07

13.

m

-Xylene

27.147 10.85

14.

p

-Xylene

27.147 2.53

15.

o

-Xylene

27.927 1.63

Peaks

RT (min.)

Conc.

(µg/mL)

1. 1,1-Dichloroethene

3.586

0.07

2. 1,1,1-Trichloroethane 8.536

0.08

3. Carbon tetrachloride 9.042

0.03

4. Benzene

9.787

0.02

5. 1,2-Dichloroethane 10.112

0.04

* DMSO interference

Column: Rxi

®

-624Sil MS, 30 m, 0.32 mm ID, 1.80 µm

(cat.# 13870)

Sample: Residual Solvents - Class 1 (cat.# 36279)

Complete analytical conditions for chromatogram

GC_PH1160 available at

www.restek.com

DB-624

Class 1 solvents coelute

Rxi®-624Sil MS

• Baseline resolution

• Easy, reliable quantification

Figure 2

The Rxi®-624Sil MS column provides complete resolution of the USP <467> Class 1 solution components—a result not often

achieved on other G43 columns.

GC_PH1160

Better resolution improves

system suitability pass rates.

When running USP <467> by headspace,

using a smaller bore liner (1 mm) can

improve system suitability pass rates.

Larger bore liners (4 mm) are used with

direct liquid injection because the sample

is vaporized in the injection port and the

liner must be able to accommodate the

solvent expansion volume. In contrast, in

headspace analysis, the sample is vapor-

ized in a vial instead of the injection port,

so a large volume liner is not needed, and

in fact it can be deleterious. In headspace

methods, using a smaller bore liner

reduces band broadening by increasing

linear velocity, allowing faster sample

transfer and improving resolution.

See p. 9 for select 1mm liners.

TECH TIP!

1. Acetonitrile

2. Dichloromethane

GC_PH00912

Resolution passes USP <467> criteria when

using a 1mm liner (red line), but fails if a

4mm liner is used (black line).

Restek carries a full line

of headspace essentials

including screw-thread headspace

vials & magnetic screw-thread

caps, Hot Swap column nuts,

1mm liners, septa, and more!

Visit

www.restek.com/usp467

for more products and tech tips

Tim Herring, Technical Service Specialist

Rxi®-624Sil MS

min.

www.restek.com