Chromatography Columns 2013+ - page 160

103
GCCOLUMNS |
APPLICATION-SPECIFICCOLUMNS
Environmental
VolatileOrganicsAnalysis
similar
phases
DB-624, HP-624,VF-624, BP-624, ZB-624,
AT-624, 007-1301, G43R
Rxi®-624SilMSColumns
(fusedsilica)
(midpolarityCrossbond® silarylenephase; equivalent to6% cyanopropylphenyl/
94%dimethyl arylenepolysiloxane)
• Lowbleed, high thermal stability column—maximum temperatures up to320 °C.
• Inert—excellent peak shape for awide range of compounds, including acidic and
basic compounds.
• Selective—highly selective for residual solvents, great choice forUSP<467>.
• Manufactured for column-to-column reproducibility—well-suited for validated
methods.
Rxi®-624SilMSStructure
X
Y
Z
Si
O
R
CH
3
CH
3
CH
3
C N
CH
3
Si
O
O
CH
3
CH
3
Si
CH
3
CH
3
Si
VolatilesbyEPAMethod8260onRxi®-624SilMS (30m,0.25mm ID,1.40μm)
Resolutionof critical pairs, low
bleed, andhigh inertnessmake
thisagreat column forvolatiles!
Column:
Rxi
®
-624Sil MS, 30m, 0.25mm ID, 1.40µm (cat.# 13868)
Sample:
8260A SurrogateMix (cat.# 30240)
8260A Internal StandardMix (cat.# 30241)
8260BMegaMix
®
CalibrationMix (cat.# 30633)
VOA CalibrationMix#1 (ketones) (cat.# 30006)
8260BAcetateMix (revised) (cat.# 30489)
CaliforniaOxygenatesMix (cat.# 30465)
502.2 CalibrationMix#1 (gases) (cat.# 30042)
Conc.:
25 ppb in ROwater
Injection
purge and trap split (split ratio 30:1)
Inj.Temp.:
225 °C
Purge and Trap
Instrument:
OI Analytical 4660
TrapType:
10Trap
Purge:
11min.@ 20 °C
Desorb PreheatTemp.: 180 °C
Desorb:
0.5min.@ 190 °C
Bake:
5min.@ 210 °C
Interface Connection: injection port
Oven
OvenTemp:
35 °C (hold5min.) to60 °Cat 11 °C/min. to220 °C
at 20 °C/min. (hold 2min.)
Carrier Gas:
He, constant flow
FlowRate:
1.0mL/min.
Detector:
MS
Mode:
Scan
Transfer LineTemp.:
230 °C
AnalyzerType:
Quadrupole
SourceTemp.:
230 °C
QuadTemp.:
150 °C
Electron Energy:
70 eV
Solvent DelayTime:
1.5min.
TuneType:
BFB
IonizationMode:
EI
Scan Range:
36-260 amu
Instrument:
Agilent 7890AGC& 5975CMSD
XIC—=m/z43
—=m/z72
GC_EV1169
For peak list and complete conditions, seepage547.
NEW!
ID
df
temp. limits
20-Meter
30-Meter
60-Meter
0.18mm 1.00µm -20 to 300/320°C 13865
0.25mm 1.40µm -20 to 300/320°C
13868
0.32mm 1.80µm -20 to 300/320°C
13870
13872
0.53mm 3.00µm -20 to 280/300°C
13871
for