Chromatography Columns 2013+ - page 137

80
GCCOLUMNS |
APPLICATION-SPECIFICCOLUMNS
Petroleum & Petrochemical
PetroleumoxygenatesonanRt®-TCEPcolumn.
similar
phases
SPB-TCEP, CP-TCEP
1
2
3
4
5
6
7
8
9
Column:
Rt
®
-TCEP, 60m, 0.25mm ID, 0.4µm (cat.# 10999)
Inj.:
1.0µL split injection, components@ 500ppm.
Oven temp.:
60°C (hold 5min.) to 100°C@ 5°C/min. (hold 10min.)
Inj./det. temp.:
200°C
Carrier gas:
helium
Linear velocity:
30cm/sec. set@ 80°C
FID sensitivity:
6.4 x 10
-11
AFS
Split flow:
46mL/min.
1. methyl
tert
-butyl ether
2.
n
-undecane
3.
tert
-butanol
4. methanol
5. isopropyl alcohol
6. ethanol
7.
n
-dodecane
8.
n
-tridecane
9.
n
-butanol
GC_PC00195
min.
2
4
6
8
10
12
14
16
Aromatics&Oxygenates inGasolineAnalysis
Rt®-TCEPColumns
(fusedsilica)
(highlypolar phase; 1,2,3-tris[2-cyanoethoxy]propane—not bonded)
• General purpose columns, ideal for aromatics andoxygenates in gasoline.
• Temperature range: 0 °C to135 °C.
Most gasolines contain aliphatic hydrocarbons up to
n
-dodecane (C12). To improve
identificationof the aromatics andoxygenates, it is desirable to elutebenzene afterC11
and toluene after C12. The extremely polar Rt®-TCEP stationary phase provides a
retention index for benzene greater than 1100 and permits the separation of alcohols
and aromatics from the aliphatic constituents in gasoline.
Rt®-TCEP columnshave the samehighpolarityasTCEPpacked columns (precolumns
inASTMMethodD4815 for the analysis of petroleumoxygenates),with the efficiency
of a capillary column. The result is a column that can separate awide variety of com-
poundswith an elutionpatternunattainableusingother highpolarity siloxanes.
TheRt®-TCEP column incorporates anonbonded stationaryphase coatedona surface
specialized for enhancedpolymer stability and extended column lifetime. Solvent rins-
ing should be avoided. Conditioning is necessary only if the column is to be used at
temperatures near themaximumoperating temperature.
SolutionsForYourPetroleum&PetrochemicalAnalyses
Visit us at
Improvedbest-in-classGCcolumns • Standards • Industryexpertsat your service.
ID
df
temp. limits
30-Meter
60-Meter
0.25mm 0.40µm 0 to 135°C 10998
10999
Website :
E-mail :
TelNo : 03 9762 2034 . . . inAUSTRALIA
Mar 2011