TCDDs on Rtx®-Dioxin2 Column

GC_EV1239

Column

Rtx®-Dioxin2, 60 m, 0.25 mm ID, 0.25 µm (cat.# 10758)

Sample

Laboratory prepared test mix

Injection

Inj. Vol.:

1.0 µL splitless

Inj. Temp.:

290 °C

Oven

Oven Temp.:

150 °C (hold 1 min) to 210 °C at 30 °C/min (hold 1 min) to 250 °C at 3 °C/min

(hold 12 min) to 330 °C at 70 °C/min (hold 6 min)

Carrier Gas

He, constant flow

Flow Rate:

1.5 mL/min

Detector

Micromass Autospec Ultima

Transfer Line

Temp.:

290 °C

Analyzer Type:

Magnetic sector

Source Temp.:

290 °C

Tune Type:

PFK

Ionization Mode: EI

Instrument

Agilent/HP6890 GC

Acknowledgement

Courtesy of Maxxam Analytics (Ontario, Canada).

GC_EV1252

GC COLUMNS |

APPLICATION-SPECIFIC COLUMNS

Environmental

Rtx®-Dioxin2 Columns

(fused silica)

(proprietary Crossbond® phase)

• Isomer specificity for 2,3,7,8-TCDD and 2,3,7,8-TCDF achieved with one

GC column.

• Thermally stable to 340 °C for longer lifetime.

• Unique selectivity for toxic dioxin and furan congeners allows use as a

confirmation GC column.

Dioxin & Furan Analysis

Restek innovation!

Excellent for dioxins or furans.

TCDFs on Rtx®-Dioxin2 Column

Column

Rtx®-Dioxin2, 60 m, 0.25 mm ID, 0.25 µm (cat.# 10758)

Sample

Laboratory prepared test mix

Injection

Inj. Vol.:

1.0 µL splitless

Inj. Temp.:

290 °C

Oven

Oven Temp.:

180 °C (hold 1 min) to 235 °C at 45 °C/min (hold 1 min) to 250 °C at 3 °C/min

(hold 15 min) to 300 °C at 50 °C/min (hold 1 min)

Carrier Gas

He, constant flow

Flow Rate:

1.7 mL/min

Detector

Micromass Autospec Ultima

Transfer Line

Temp.:

290 °C

Analyzer Type:

Magnetic sector

Source Temp.:

290 °C

Tune Type:

PFK

Ionization Mode: EI

Instrument

Agilent/HP6890 GC

Acknowledgement

Courtesy of Maxxam Analytics (Ontario, Canada).

“Using the Rtx®-Dioxin2 column

allowed us to combine EPA 1613

TCDD-only and TCDF confirmation

analyses onto one column and one

instrument. This resulted in multiple

benefits—we shortened run times,

reduced instrument downtime and

column changes, and increased instru-

ment capacity for our full list samples.”

Owen Cosby

Supervisor, HRMS Services

Maxxam Analytics

40-Meter

60-Meter

ID df

temp. limits

cat.#

price

cat.#

price

0.18 mm 0.18 µm 20 to 320/340 °C

10759

0.25 mm 0.25 µm 20 to 320/340 °C

10758