GC COLUMNS |
GENERAL-PURPOSE COLUMNS
GCxGC
Figure 1:
In a contour plot like this one showing clear determination of over 80 pesticides in marijuana, the x-axis represents the
primary column retention time and the y-axis represents the secondary column retention time. Peaks aligned along the y-axis
would coelute in one-dimensional GC, which is especially problematic if they cannot then be separated by MS.
GC_FF1205
Column:
Rxi®-5Sil MS 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623); Rtx®-200 1.3 m, 0.25 mm ID, 0.25 µm (cat.# 15020);
Sample:
Diluent: Toluene;
Injection:
Inj. Vol.: 1 µL splitless (hold 1 min); Liner:
Sky™ 4 mm single taper w/wool (cat.# 23303.1); Inj. Temp.: 250 °C; Purge Flow: 40 mL/min;
Oven:
Oven Temp.: Rxi®-5Sil MS: 80 °C (hold 1 min) to 310 °C at 5 °C/min; Rtx®-200: 85 °C (hold 1 min)
to 315 °C at 5 °C/min;
Carrier Gas:
He, corrected constant flow (2 mL/min);
Modulation:
Modulator Temp. Offset: 20 °C; Second Dimension Separation Time: 3 sec; Hot Pulse Time: 0.9 sec; Cool Time
between Stages: 0.6 sec;
Detector:
TOFMS; Transfer Line Temp.: 290 °C; Analyzer Type: TOF; Source Temp.: 225 °C; Electron Energy: 70 eV; Mass Defect: -20 mu/100 u; Solvent Delay Time: 5 min; Tune
Type: PFTBA; Ionization Mode: EI; Acquisition Range: 45-550 amu; Spectral Acquisition Rate: 100 spectra/sec;
Instrument:
LECO Pegasus 4D GCxGC-TOFMS;
Notes:
Rtx®-200 (cat.# 15020) is a 15 m
column. A 1.3 m section was used as the second dimension column.
For a peak list, visit
www.restek.comand enter chromatogram GC_FF1205 in the search function.
Coelution in
one-dimensional GC!
Check out the Restek blog for the most current topics in GCxGC.
blog. restek.comTo choose the perfect
primary/secondary
column combination for
your application, use our
guide on
page 60!