GC COLUMNS |

GENERAL-PURPOSE COLUMNS

GCxGC

Figure 1:

In a contour plot like this one showing clear determination of over 80 pesticides in marijuana, the x-axis represents the

primary column retention time and the y-axis represents the secondary column retention time. Peaks aligned along the y-axis

would coelute in one-dimensional GC, which is especially problematic if they cannot then be separated by MS.

GC_FF1205

Column:

Rxi®-5Sil MS 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623); Rtx®-200 1.3 m, 0.25 mm ID, 0.25 µm (cat.# 15020);

Sample:

Diluent: Toluene;

Injection:

Inj. Vol.: 1 µL splitless (hold 1 min); Liner:

Sky™ 4 mm single taper w/wool (cat.# 23303.1); Inj. Temp.: 250 °C; Purge Flow: 40 mL/min;

Oven:

Oven Temp.: Rxi®-5Sil MS: 80 °C (hold 1 min) to 310 °C at 5 °C/min; Rtx®-200: 85 °C (hold 1 min)

to 315 °C at 5 °C/min;

Carrier Gas:

He, corrected constant flow (2 mL/min);

Modulation:

Modulator Temp. Offset: 20 °C; Second Dimension Separation Time: 3 sec; Hot Pulse Time: 0.9 sec; Cool Time

between Stages: 0.6 sec;

Detector:

TOFMS; Transfer Line Temp.: 290 °C; Analyzer Type: TOF; Source Temp.: 225 °C; Electron Energy: 70 eV; Mass Defect: -20 mu/100 u; Solvent Delay Time: 5 min; Tune

Type: PFTBA; Ionization Mode: EI; Acquisition Range: 45-550 amu; Spectral Acquisition Rate: 100 spectra/sec;

Instrument:

LECO Pegasus 4D GCxGC-TOFMS;

Notes:

Rtx®-200 (cat.# 15020) is a 15 m

column. A 1.3 m section was used as the second dimension column.

For a peak list, visit

www.restek.com

and enter chromatogram GC_FF1205 in the search function.

Coelution in

one-dimensional GC!

Check out the Restek blog for the most current topics in GCxGC.

blog. restek.com

To choose the perfect

primary/secondary

column combination for

your application, use our

guide on

page 60

!