GC COLUMNS |

PLOT COLUMNS

Fused Silica Porous Polymer Columns, Particle Trap

PLOT Column Particle Trap

• Includes two Press-Tight® connectors and a 2.5 m column.

• Protects detector and valves; connects between column and detector or valve.

• Eliminates detector spikes and scratches in valve rotors.

The technology used to adhere particles in PLOT columns is excellent; however, it

is still possibile for particles to dislodge when extreme pressure shocks and gas flow

changes occur. This sometimes happens when valve switching or backflushing is used.

In those cases, using particle traps is recommended.

Description

qty.

cat.#

price

PLOT Column Particle Trap, 2.5 m, 0.32 mm ID with 2 Press-Tight Connectors

ea.

19753

PLOT Column Particle Trap, 2.5 m, 0.53 mm ID with 2 Press-Tight Connectors

ea.

19754

19754

Rt®-U-BOND Columns

(fused silica PLOT)

divinylbenzene ethylene glycol/dimethylacrylate

• Restek's highest polarity porous polymer column.

• Polar PLOT column, incorporating divinylbenzene ethylene glycol/

dimethylacrylate.

• Highly inert for the analysis of polar and nonpolar compounds.

• Stable to 190 °C.

similar

phases

HP-PLOT U, CP-PoraPLOT U, CP-PoraBOND U

15-Meter

30-Meter

ID

df

temp. limits

cat.#

price

cat.#

price

0.25 mm 8 µm to 190 °C

19771

19772

0.25 mm 12 µm to 190 °C

19782

0.32 mm 10 µm to 190 °C

19751

19752

0.53 mm 20 µm to 190 °C

19749

19750

Formaldehyde on Rt®-U-BOND

GC_CH01137

Column

Rt®-U-BOND, 30 m, 0.53 mm ID, 20 µm (cat.# 19750)

Sample

Formaldehyde (manual headspace)

Injection

Inj. Vol.:

10 µL split (split ratio 10:1)

Liner:

2 mm split Precision® liner w/wool (cat.# 20823)

Inj. Temp.:

200 °C

Split Vent

Flow Rate:

40 mL/min

Oven

Oven Temp.:

100 °C (hold 1 min) to 150 °C at 25 °C/min (hold 3 min)

Carrier Gas

He, constant pressure (7.7 psi, 53.1 kPa)

Linear Velocity:

39 cm/sec @ 100 °C

Detector

TCD@ 200 °C

Make-up Gas

Type:

He

Data Rate:

20 Hz

Sensitivity Mode: He/H

2

Instrument

HP5890 GC

Peaks

1. Air

2. Carbon dioxide

3. Formaldehyde

4. Water

5. Methanol

Excellent peak shape for highly polar

analytes!