70

www.restek.com

GC COLUMNS |

APPLICATION-SPECIFIC COLUMNS

Environmental

Rtx®-PCB Columns

(fused silica)

(proprietary Crossbond® phase)

• Unique polymer for PCBs analysis by GC-ECD or GC-MS.

• Good results for other semivolatiles.

• Low polarity; inert to active compounds.

• Stable to 340 °C.

*Maximum temperatures listed are for shorter length columns. Longer columns may have a different maximum temperature.

PCB Congeners Analysis

Restek innovation!

Aroclor PCBs on Rtx®-PCB

6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00 30.00 32.00 34.00 36.00 38.00 40.00 42.00 44.00 46.00 48.00 50.00 52.00 54.00 56.00 58.00 60.00 62.00 64.00

28

52

101

118

153

138

180

Time (min)

GC_EV00767

Column

Rtx®-PCB, 60 m, 0.25 mm ID, 0.25 µm

(cat.# 13226)

Sample

Aroclor 1242 (cat.# 32009)

Aroclor 1254 (cat.# 32011)

Aroclor 1262 (cat.# 32409)

Conc.:

333 ppm each

Injection

Inj. Vol.:

1.0 µL splitless (hold 0.75 min)

Liner:

Splitless taper (4 mm) w/wool

(cat.# 22405)

Inj. Temp.:

280 °C

Oven

Oven Temp.:

100 °C (hold 1 min) to 200 °C at 30 °C/

min to 320 °C at 2 °C/min (hold 1 min)

Carrier Gas

He, constant flow

Flow Rate:

1.1 mL/min

Detector

MS

Mode:

Scan

Transfer Line

Temp.:

280 °C

Ionization Mode: EI

Scan Range:

50-550 amu

European indicator PCB

congeners resolved from other

congeners when using GC-MS.

Peaks identified with PCB congener numbers.

20-Meter

30-Meter

40-Meter

60-Meter

ID

df

temp. limits*

cat.#

price

cat.#

price

cat.#

price

cat.#

price

0.18 mm 0.18 µm 30 °C to 320 °C 41302

41303

41304

0.25 mm 0.25 µm 30 °C to 320/340 °C

13223

13226

0.32 mm 0.50 µm 30 °C to 320/340 °C

13239

See

pages 193–202 .

Satisfaction