Carbamates on Ultra Carbamate

LC COLUMNS

Application-Specific Columns

Column Characteristics:

particle size:

3 µm or 5 µm, spherical

pore size:

100 Å

pH range:

2.5 to 8

temperature limit: 80 °C

Column

Ultra Carbamate (cat.# 9177352)

Dimensions:

50 mm x 2.1 mm ID

Particle Size:

3 µm

Pore Size:

100 Å

Temp.:

ambient

Sample

531.1 Carbamate Pesticide Calibration Mixture (cat.# 32273)

4-bromo-3,5-dimethylphenyl-N-methylcarbamate (BDMC)

(cat.# 32274)

Diluent:

methanol

Conc.:

50 µg/mL

Inj. Vol.:

1 µL

Mobile Phase

A:

2 mM ammonium acetate:methanol (v/v, 90/10)

B:

2 mM ammonium acetate:methanol (v/v, 10/90)

Time (min) %A %B

0.00

80 20

20

0 100

25

0 100

Flow:

0.2 mL/min

Detector

LECO Unique® TOFMS

Run Length:

25 min

Ionization Source Type: high flow ESI

Ion Mode:

positive

Desolvation Temp.:

130 °C

Nebulizing Pressure:

100 kPa

Desolvation Gas (N

2

):

4 L/min

Interface Temp.:

120 °C

Nozzle Voltage:

62 V

Capillary Voltage:

2.75 kV

Instrument

Agilent 1100

Acknowledgement

LECO Corporation

0 2.5

1

2

3

4

5

6

7

8

9

10

11

5.0 7.5 10.0 12.5 15.0 17.5 20.0 Time (min)

LC_FF0473

Peaks

1. Aldicarb sulfone

2. Aldicarb sulfoxide

3. Oxamyl

4. Methomyl

5. 3-Hydroxycarbofuran

6. Aldicarb

7. Propoxur

8. Carbofurane

9. Carbaryl

10. Methiocarb

11. BDMC (IS)

Chromatographic Properties

Restek chemists developed the Ultra Carbamate column specifically for carbamates analysis.

The unique packing separates 10 target carbamates in just over 10 minutes. The column is

compatible with fluorescence or LC-MS detection.* An Ultra Carbamate column can process

as many as three samples per hour, versus less than two samples per hour on a general-purpose

C18 column. In addition to increased sample throughput, this much faster analysis will

significantly reduce solvent usage—and the costs of disposing of solvent waste.

Ultra Carbamate Columns

*For post-column derivatization/fluorescence detection applications using a 4.6 mm ID column, the total system dead

volume, including the post-column reactor, must be less than 650 µL. For standard post-column reactor systems, we

recommend a 250 mm x 4.6 mm, 5 µm column. Contact Restek® Technical Service or your local Restek® representative

for more information.

Ultra Carbamate Guard Cartridges

Guard Cartridges

3-pk.

(10 x 2.1 mm)

3-pk.

(10 x 4.0 mm)

price

Ultra Carbamate Guard Cartridge

917750212

917750210

2.1 mm ID

3.0 mm ID

4.0 mm ID

4.6 mm ID

Length

cat.#

price

cat.#

price

cat.#

price

cat.#

price

3 µm Columns

50 mm 9177352

917735E

9177354

9177355

100 mm 9177312

917731E

9177315

5 µm Columns

250 mm

9177575