Table of Contents Table of Contents
Previous Page  104 / 629 Next Page
Information
Show Menu
Previous Page 104 / 629 Next Page
Page Background

102

www.restek.com

GC COLUMNS |

APPLICATION-SPECIFIC COLUMNS

Specially Deactivated

Basic Compounds Analysis

Stabilwax®-DB Columns

(fused silica)

(polar phase; Crossbond® base-deactivated Carbowax® polyethylene glycol—for amines and

basic compounds)

• Application-specific columns for underivatized amines and other basic compounds,

including alkylamines, diamines, triamines, nitrogen-containing heterocyclics. No

need for column priming.

• Temperature range: 40 °C to 220 °C.

Stabilwax®-DB columns reduce adsorption and improve responses for many basic

compounds, without analyte derivatization or column priming. For different selectiv-

ity of basic compounds, or higher oven temperatures, use an Rtx®-5 Amine column.

Stabilwax®-DB is a bonded stationary phase, but avoid rinsing these columns with

water or alcohols.

similar phases

CAM, CP-WAX 51 for Amines,

Carbowax® Amine

15-Meter

30-Meter

60-Meter

ID

df

temp. limits

cat.#

price

cat.#

price

cat.#

price

0.25 mm 0.25 µm 40 to 210/220 °C 10820

10823

0.50 µm 40 to 210/220 °C

10838

0.32 mm 0.25 µm 40 to 210/220 °C 10821

10824

0.50 µm 40 to 210/220 °C

10839

1.00 µm 40 to 210/220 °C 10851

10854

10857

0.53 mm 0.50 µm 40 to 210/220 °C

10840

1.00 µm 40 to 210/220 °C 10852

10855

10858

1.50 µm 40 to 210/220 °C

10869

Volatile Organic Compounds by U.S. EPA Method 1671 on Stabilwax®-DB

GC_EV00793

Columns

Stabilwax®-DB 30 m, 0.32 mm ID, 1.00 µm

(cat.# 10854)

and Stabilwax® 30 m, 0.32 mm ID, 1.00 µm

(cat.# 10654)

using IP deactivated guard column 5 m, 0.53 mm ID

(cat.# 10045)

with SeCure® “Y” connector kit (cat.# 20278)

Sample

1671 Volatile organics mix

Diluent:

Deionized water

Injection

Inj. Vol.:

1.0 µL split (split ratio 12:1)

Liner:

Gooseneck splitless (4 mm) (cat.# 20798)

Inj. Temp.:

200 °C

Oven

Oven Temp.:

40 °C (hold 5 min) to 180 °C at 7 °C/min (hold 5 min)

Carrier Gas

He, constant pressure

Linear Velocity:

39.68 cm/sec @ 40 °C

Detector

FID@ 250 °C

Notes

“Y” Press-Tight® Connector (cat.# 20405) also used

Column 1 (Stabilwax®) linear velocity: 39.25 cm/sec

@ 40 °C

Peaks

t

R

(min)

Conc. (µg/mL)

1. Dimethylamine

1.71

200

2. Methylamine

1.76

200

3. Diethylamine

2.46

200

4. Triethylamine

2.64

200

5. Tetrahydrofuran (IS)

4.88

100

6. Methanol

7.12

40

7. Ethanol

8.01

40

8. Acetonitrile

9.82

200

9.

n

-Propanol

11.03

200

10. Methyl Cellosolve®

15.56

200

11. Formamide

18.68

500

12. Dimethyl sulfoxide

23.75

100

13. Ethylene glycol*

500

*Included in sample, but does not elute due to base deactivation in

the DB phase.