1. DISMOUNTING/ MOUNTING OF THE PISTON:
1.1 Dismounting:
1.2 Unlock safety turn-lock fastener by turning clockwise.
1.3 Swing open the divisible column lock at the sealing joint.
1.4 Remove cartridge from protective jacket by pressing fixed piston.
CAUTION! - dismounting of the end nut is not necessary.
1.5 Mounting
1.6 Insert cartridge into protective jacket to stopper.
CAUTION! - Pay attention to safety turn-lock fastener when cartridge is inserted! Posi-
tion the piston square correspondingly.
1.7 Close divisible column lock.
1.8 Lock safely turn-lock fastener by turning it clockwise.
2. FRIT REPLACEMENT:
2.1 Eject old frit with supplied frit ejector.
2.2 Insertion of new frit: The teflon® receptacle of the piston is protected when it is heated
before the new frit is pressed in (max. 121º C).
3. MOUNTING OF MPLC-FITTINGS:
These mounting instructions are for tubing with 1.6 and 3.2 mm outer diameter.