Multiple Gas Analyzer + Sulfur GC Systems
16
|
|
|
|
8610-0073
Detect low sulfur
concentrations
Multiple gas analysis
AND sulfur analysis in
one unit
Uses just 1 gas
sampling valve and 3
columns
Built-in, "whisper quiet"
air compressor
6 channel PeakSimple
data system
...on the compact
chassis, or on
the 8610V chassis for
just $2,000 more.
The inert
sample
loop is
visible
on the
front of
the
valve
oven.
Many analysts require natural gas analysis for various applications. Frequently, sulfur compounds are also of interest.
Because low sulfur concentrations (<50ppm) are difficult to measure, SRI has enhanced our popular
to simultaneously monitor low levels of sulfur compounds. The additional hardware required is an
, which selectively detects sulfur down to mid-ppb range, and a room temperature sulfur-inert sample loop.
One reason sulfur is so difficult to measure is that it disappears on contact with hot stainless steel surfaces; even
limited contact with a hot stainless steel loop will destroy any sulfur in the gas sample. Our solution is to use an inert
transfer line leading to a splitter, and a 60-meter thick film capillary column. While Teflon columns are sometimes also
used for sulfur analysis, the natural gas analysis (using molecular sieve and HayeSep D) requires column temperatures
of 250
°
C or higher. Since the sulfur column is located in the column oven, it is essential to use a column like the 60-
meter capillary which can tolerate the higher temperatures.