Restek 2011/12 Chromatography Catalog - page 643

643
GC CHROMATOGRAMS |
PERSONAL CARE & CLEANING PRODUCTS
Cleaning Solvents
Cleaning Solvents (All-Purpose Cleaner)
Rtx®-VMS
Column:
Rtx
®
-VMS, 60m, 0.25mm ID, 1.4µm (cat.# 19916)
Sample:
approx. 5% cleaner in methanol
Inj.:
1.0µL split (split ratio 100:1), 4mm inlet liner (cat.# 20814)
Inj. temp.:
250°C
Carrier gas:
hydrogen, constant flow
Flow rate:
1.5mL/min.
Oven temp.:
60°C (1 min. hold) to 138°C @ 4°C/min. to 145°C @ 2°C/min. to 210°C
@ 15°C/min. to 230°C @ 5°C/min.
Det.:
FID @ 250°C
1. butyl Cellosolve
®
2. limonene
GC_FF00646
Cleaning Solvents (Glass Cleaner)
Rtx®-VMS
Column:
Rtx
®
-VMS, 60m, 0.25mm ID, 1.4µm (cat.# 19916)
Sample:
approx. 5% cleaner in methanol
Inj.:
1.0µL split (split ratio 100:1), 4mm inlet liner (cat.# 20814)
Inj. temp.:
250°C
Carrier gas:
hydrogen, constant flow
Flow rate:
1.5mL/min.
Oven temp.:
60°C (1 min. hold) to 138°C @ 4°C/min. to 145°C @ 2°C/min. to 210°C
@ 15°C/min. to 230°C @ 5°C/min.
Det.:
FID @ 250°C
1.
iso
-propanol
2. butyl Cellosolve
®
GC_FF00647