METERS
Temperature 4,8 Channels Dataloggers
TM-747D, Datalogger Thermometer 4 Channels Include Software/
Cable
The TM-747D is a rugged, easy-to-use thermometer with 4 mini, flat-pin,
thermocouple sockets for inputs. It accepts 7 different thermocouple types
and displays all 4 inputs at the same time. It also provides differential
temperature measurement readings of T3-T4, as well as individual readings
of the 4 inputs. Other features are a low battery indicator, perpetual
calendar, and auto power-off.
Multi function for K/J/E/T/R/S/N types.
4 Backlit Displays.
Resolution 0.1°C/0.1°F, 1°C/1°F.
Auto Power-Off, Low Battery Indication.
USB Interface with Windows Software (Optional RS232 Cable Available).
10,000 Record Datalogger.
CALL-Fit to Quick- Read Memory Data (50 Pages/Second).
Perpetual Calendar Function.
Input protection at thermocouple input: 24V AC/DC Maximum.
Dimension (Includes housing) (mm): L184 x W82 x H44.
Weight (Includes housing, batteries): Approx. 450g.
Battery Life: 4 "AAA"" batteries (included), 550 hr.
TM-747D
PROVA 800, 8 Input Thermometer / Datalogger
8 input temperature measurement/logging °C/°F)
Graphic display of
each input
11 types of thermocouple (K, J, E, T, R, S, N, L, U, B, C)
Easy
thermocouple mini connector
Basic accuracy 0.05% ±1°C (Ktype)
Sampling Rate: 1 sec./8 input
Programmable Hi-La alarm for 8 inputs
Display of max and min values of 8 inputs
Standard 2G SD memory card
(stores 3.8-year data)
Optional 8G SDHC memory card
SD card directly
read by PC
Easy data file management (Read and Delete functions)
Built-in calendar clock
Independent Input Setup (type of thermocouple,
Hi-Lo alarm values, name of Engineering Unit)
Isolated Input Protection of
350Vp-p between any two inputs
Programmable Engineering Unit (E.U.) to
integrate Analog Output (-50 to 50mV) from instruments
Screen hardcopy
into files
Paperless Recorder.
PROVA 800
Sensor
Type
°
C
°
F
Range
Accuracy
Range
Accuracy
K
-200 ~ -100
-100 ~ -1370
±2.0°C
±0.05% ±1.0°C
-328 ~ -148
-148 ~ -2498
±3.6°F
±0.05% ±1.8°F
J
-200 ~ -100
-100 ~ 100
100 ~ 1000
±2.5°C
±1.5°C
±0.05% ±1.0°C
-328 ~ -148
-148 ~ 212
212 ~ 1832
±4.5°F
±2.7°F
±0.05% ±1.8°F
E
-150 ~ -100
-100 ~ 760
±2.0°C
±0.05% ±1.0°C
-238 ~ -148
-148 ~ 1400
±3.6°F
±0.05% ±1.8°F
T
-200 ~ -100
-100 ~ 400
±1.5°C
±0.1% ±0.5°C
-328 ~ -148
-148 ~ 752
±2.7°F
±0.1% ±0.9°F
R/S
0 ~ 100
100 ~ 300
300 ~ 1600
±5.0°C
±3.0°C
±0.05% ±2.0°C
32 ~ 212
212 ~ 572
572 ~ 2912
±9.0°F
±5.4°F
±0.05% ±3.6°F
N
0 ~ 1300
±0.1% ±1.0°C
32 ~ 2372
±0.1% ±1.8°F
L
-200 ~ 900
±0.1% ±1.0°C
-328 ~ 1652
±0.1% ±1.8°F
U
0 ~ 600
±0.1% ±1.0°C
32 ~ 1112
±0.1% ±1.8°F
B
600 ~ 1820
±0.05% ±2.0°C
1112 ~ 3308
±0.05% ±3.6°F
C
0 ~ 2310
±0.1% ±1.5°C
32 ~ 4190
±0.1% ±2.7°F
mV
-50 ~ 50
±0.05% ±20µV
-50 ~ 50
±0.05% ±20µV
Sensor
Type
°
C
°
F
Thermocouple Range
Accuracy
(18 to 28°C Ambient)
Thermocouple Range
Accuracy
(18 to 28°C Ambient)
K
-100 ~ 1300
±0.1% rdg + 0.7°C
-148 ~ -2372
±0.1% rdg + 1.4°F
J
-100 ~ 1000
±0.1% rdg + 0.7°C
-148 ~ 1832
±0.1% rdg + 1.4°F
E
-50 ~ 800
±0.1% rdg + 0.7°C
-58 ~ 1472
±0.1% rdg + 1.4°F
T
-100 ~ 400
±0.1% rdg + 0.7°C
-148 ~ 752
±0.1% rdg + 1.4°F
R/S
0 ~ 1700
± 0.1% rdg + 2°C
32 ~ 3092
±0.1% rdg + 4°F
N
-100 ~ 1300
± 0.1% rdg + 1.5°C
-148 ~ 2372
± 0.1% rdg + 3°F
Operating Temperature and Humidity: 0 to 50°C (32 to 122°F) 0 to 80% RH
Storage Temperature and Humidity: -20 to 60°C (-4 to 140°F) 0 to 80% RH