![Show Menu](styles/mobile-menu.png)
![Page Background](./../common/page-substrates/page0024.jpg)
DETECTOR TYPE
Flame Ionisation Detector
RANGE
0–10 ppm
0–100 ppm
0–1000 ppm
other range possible
REPEATABILITY
< 1% full scale
ACCURACY
Better than ±1% full scale
STANDARD FEATURES
• Manual or autoranging (user selectable)
• Microprocessor controlled
• 5.6” TFT intelligent LCD module with Touch Screen
• Self diagnosis system with auto-resolve alarm
• 4-20 mA isolated output
• Alarm Historic
• Digital ouputs for remote monitoring: (all dry relay contacts)
- System status (1 output)
- Range in use (3 output)
- Calibration in use (1 output)
OPTIONS
• Serial port: RS-232 / 422 / 485 / Profibus
• 2 alarm outputs (user programmable set point)
SAMPLE FLOW REQUIREMENTS
50 to 200 sccm
AIR FLOW REQUIREMENTS
200 to 600 sccm
FUEL FLOW REQUIREMENTS
40 to 150 sccm
RECOMMENDED MAXIMUM
OPERATING PRESSURE
RECOMMENDED MINIMUM
OPERATING PRESSURE
OPERATING TEMPERATURE
40 PSIG
10 PSIG
10
°
C to 45
°
C
SUPPLY
115 VAC
50 – 60 Hz or 220 VAC
50 – 60 Hz
DRIFT
< ± 1% over 24 hours
WEIGHT
27 lbs (12 kg)
POWER CONSUMPTION
Maximum 50 watts
GAS CONNECTIONS
Sample: 1/8’’ compression fittings
Vent: 1/8’’ compression fitting
CALIBRATION GAS
Zero: LDP1000 purified gas (Getter)
Span: 80%-90% of full scale of methane
SPECIFICATIONS:
APPLICATIONS:
• Air separation unit
• Cryogenic truck loading station
• Speciality gas laboratories
• Process control
• Steel Industries
• Chemical plants
• Welding gas control
• Gas management system
• Quality control for truck fills and gas cylinders
• Safety
• Product validation
• Scrubber & oxidizer efficiency
• Carbon bed breakthrough detection
• Well logging
• Industrial hygiene & safety monitoring
• Fence line (perimeter) monitoring around industrial sites
• CEMS (continuous emission monitoring systems)