Background Image
Table of Contents Table of Contents
Previous Page  4 / 12 Next Page
Basic version Information
Show Menu
Previous Page 4 / 12 Next Page
Page Background

Figure 1:

QuEChERS dSPE cleanup removes interferences that obscure target pesticides.

visit

www.restek.com/quechers

4

QuEChERS dSPE Cleanup Assures Optimal Results for Pesticide Analysis

Improves integration and mass spectral matches.

Removes matrix interferences that obscure target analytes or cause ion suppression.

Protects GC inlet, and LC and GC columns from contamination.

allethrin

buprofezin

cis

-chlordane

trans

-chlordane

chlorpyrifos

cyprodinil

dacthal

diphenamid

endosulfan I

endosulfan II

fenthion

metolachlor

myclobutanil

oxyfluorfen

pendimethalin

pentachlorothioanisole

pirimiphos methyl

triadimefon

triadimenol

Without extract cleanup, hexadecanoic acid from the

matrix obscures peaks for all the following pesticides.

(example XIC chromatogram = endosulfan I)

Q-sep™ QuEChERS tubes easily

remove matrix interferences.

Effective...

GC_FF01125-27

Column

Rxi

®

-5Sil MS, 20 m, 0.18 mm ID, 0.18 μm (cat.# 43602)

Sample

Sweet potato spiked with pesticide mix and extracted with acetoni-

trile and Q-sep™ QuEChERS EN Method 15662 extraction salts

Injection

Inj. Vol.:

1.0 μL splitless (hold 1 min)

Liner:

Gooseneck splitless (4 mm) w/deact. wool (cat.# 22405)

Inj. Temp.:

250 °C

Oven

Oven Temp: 72.5 °C (hold 1 min) to 350 °C at 20 °C/min

Carrier Gas

He, constant flow

Flow Rate:

1.2 mL/min

A. No extract cleanup

B. Q-sep™ QuEChERS cleanup

Detector

MS

Mode:

Transfer Line Temp.: 300 °C

Analyzer Type:

TOF

Ionization Mode:

EI

Acquisition Range: 45-550 amu

Instrument

Agilent/HP6890 GC

Notes

A. Extract (without cleanup step) acidified with formic acid to pH 5

B. Extract with cleanup using Q-sep™ QuEChERS dSPE cleanup

tube (cat.# 26124) acidified with formic acid to pH 5.

Scan range: m/z 60, 73, 87, 129, 256 plotted