Chromatography Columns 2013+ - page 81

24
GCCOLUMNS |
CAPILLARYCOLUMNS
Selecting a GC Column
Table I
Retention indices forRestekphases
Phase
Benzene
Butanol
Pentanone
Nitropropane
Rtx-1
651
651
667
705
Rtx-5/Rtx-5MS
667
667
689
743
Rtx-20
711
704
740
820
Rtx-1301/Rtx-624
689
729
739
816
Rtx-35
746
733
773
867
Rtx-200
738
758
884
980
Rtx-50
778
769
813
921
Rtx-1701
721
778
784
881
Rtx-65TG
794
779
825
938
Rtx-225
847
937
958
958
Stabilwax
963
1158
998
1230
For FastGC,Windows®NT, 2000, XP, Vista,
orWindows®7 (compatibilitymode).
Pro
ez
GC softwarewill saveyou time
andmoneybygreatlyenhancingyour
productivityand increasing sample
throughput.
Pro
ez
GCMethodsDevelopment
Software
•Optimize temperature and flow programswith a
single analysis.
• Reduce analysis time and improve sample
resolution.
•Model retention gap and guard column applica-
tions, includingRestek Integra-Guard
®
columns.
•Optimize dual-column run conditions, columns in
parallel or in series.
Take the guesswork out of selecting the best column
and conditions for your GC analysis. Pro
ez
GC soft-
ware accurately predicts separations on any capillary
column, and is useful for selecting a column and
conditions from a singleGC run. Using your retention
data, or the extensive library, you can automatically
evaluate thousands of combinations of column dimen-
sions, oven temperature programs, and carrier gas
pressure programs to determine the best separation
with the fastest analysis time.
Pro
ez
GC includes amaster set of retention index
libraries at no extra charge!These libraries contain
more than 3,000 compounds analyzed on themost
commonly used stationary phases, in ten application
areas, including pesticides, PCBs, dioxins/furans,
flavor and fragrance compounds, drugs of abuse,
FAMEs, semivolatile and volatile pollutants, petroleum
hydrocarbons, and solvents and chemicals.The
libraries permit computer simulationwithout entering
actual laboratory data.
Description
qty.
cat.#
price
Pro
ez
GCMethodDevelopment
Software CD-ROM
ea.
21487
$201
Table II
Typical characteristics for columnswith the samephase ratio, such
as0.10mm IDx0.10μmand0.18mm IDx0.18μm, etc.
Retention,k
Thecapacityof thecolumn relates tohowmuchmaterial canbe injectedontoacolumn
without adversely affecting peak shape. If the amount of a compound (mass) exceeds
the capacity of a column (WCOT), the peakwill front,which sometimes can look like
a“shark fin”.Thegoal is to select acolumnwith sufficient capacity such thatpeak shape
will not suffer. Peak symmetry is typically used to calculate the degree of sample over-
load. There are twoprimary column-relateddimensions that affect capacity, assuming
the proper column phase was selected: column internal diameter (ID) and phase film
thickness (µ).
When selecting column ID, consideration should include the type of injection, the
detector beingused, and the concentrationof sample (amount on-column).The injec-
tion technique is an important consideration because the ID of the columnmay need
to be selected based on whether a split, splitless, cool on-column injection, or other
sample transfer to the column is being used. The second consideration is the detector
andhowmuch flow it canoptimallyworkunder. For example, someMSdetectors can
onlyhandle column flow rates of up to 1.5mL/min.; therefore, a 0.53mm ID column,
which requireshigher flows forproper chromatography, isnot anoption for thisdetec-
tor. The third consideration is sample capacity of the column. If the concentration of
the sample exceeds the column capacity, loss of resolution, poor reproducibility, and
peak distortion will result. Table II shows several typical column characteristics for
various column IDs.
Column ID
Characteristic
0.10mm 0.15mm 0.18mm 0.25mm 0.32mm 0.53mm
Helium Flow
(@ 20cm/sec.)
0.16mL/min.
0.3mL/min.
0.3mL/min.
0.7mL/min.
1.2mL/min.
2.6mL/min.
Hydrogen Flow
(@ 40cm/sec.)
0.32mL/min.
0.6mL/min.
0.6mL/min.
1.4mL/min.
2.4mL/min.
5.2mL/min.
Sample Capacity
(max load per component)
<10ng
<40ng
<50ng
50–100ng 400–500ng 1000–2000ng
Theoretical Plates/Meter
8000
4000
3500
3200
2500
1800
Table I lists the Kovats retention indices for the more common stationary phases.
Assigning a retention index to eachprobe listedprovides a basis for comparing several
stationary phases and their relative retention to one another for a set of molecular
probes. For example,whenKovats indices are identical on two columnphases, then the
resulting separations will be identical. If, however, a Kovats value of one probe varies
significantly from the value on another phase for the same probe, then the resulting
compound elution order will differ. Thus, the Kovats indices are useful for comparing
column selectivity for different types of compounds amongdifferent phases.