TECHNICAL

INFORMATION

027

112

TEL: +1 (970) 267-5200

TOLL FREE in USA: (888) 404-5837

WWW.NORDSONMEDICAL.COM

805 West 71st Street • Loveland, CO 80538 • USA

Performance of Barb to Tube Connections -

Tensile Pull Test (lbs) - Hydrostatic Burst Test (psi)

Tensile pull and burst tests were performed by an independent laboratory under controlled conditions to provide comparative

performance data for the various barb styles and sizes offered by Nordson MEDICAL. Graph numbers 1 through 8 present the

data from each of the tests for each barb size and each barb style connected to PVC tube.

The values shown are the means of a statistically significant sample quantity of each configuration. Nordson MEDICAL barbed

fittings specifically designed for use with the tubing size indicated were used in these tests. No oversized barbs, solvents,

adhesives, or secondary fasteners of any kind were used.

Tensile Pull Test:

Tensile properties were measured to determine the force required to remove the barbed fitting from the tubing.

Tensile specimens were mounted axially in a holding fixture and tested at a rate of 20 inches per minute. Failure mode was tubing

separating from the barb.

Hydrostatic Burst Test:

Hydrostatic burst testing was performed in general accordance with ASTM D1599. Failure mode was

either tubing separating from the barb, the tubing bursting, or a combination of the two.

TECHNICAL:

Barb Performance

The data presented here is for reference only. It was compiled to provide our customers with a means of comparing the characteristics of components at the time of publication.

The particular conditions of your use and application of our products are beyond our control. Thus, it is imperative that you test our products in your specific application to

determine their suitability. All information is provided without implied or express warranty or guarantee by Nordson MEDICAL or other manufacturers. None of the information

provided constitutes a recommendation or endorsement of any kind by Nordson MEDICAL.

Graph 1: 1/16" Barb Performance

Graph 2: 3/32" Barb Performance

Graph 3: 1/8" Barb Performance

Graph 4: 5/32" Barb Performance

Graph 5: 3/16" Barb Performance

Graph 6: 1/4" Barb Performance

600 Series Leak Tests

600 Series Tensile Pull Tests

10

10

10

10

10

3/8”

1/2”

5/8”

3/4”

1”

C-Flex®

20

28

32

40

55

PVC

7

9

9

11

14

Silicone

60

50

40

30

20

10

0

121

63

106

89

49

3/8”

1/2”

5/8”

3/4”

1”

C-Flex®

295

105

179

81

80

PVC

64

49

64

43

27

Silicone

300

200

100

0

43

153

500

1

10

100

1000

Units

Barb Style

1

10

100

1000

8

363

500

7

200

4

186

400

380

Classic

Units

Barb Style

311

11

1

10

100

1000

11

154

500

19

217

200

10

111

400

22

211

Classic

Units

Barb Style

38

224

1

10

100

1000

500

48

223

200

22

134

400

52

201

Classic

Units

Barb Style

9

12

6

13

215

259

124

250

1

10

100

1000

500

200

400

Classic

Units

Barb Style

33

262

21

204

500

32

257

200

Classic

1

10

100

1000

Units

Barb Style

11