Table of Contents Table of Contents
Previous Page  41 / 629 Next Page
Information
Show Menu
Previous Page 41 / 629 Next Page
Page Background www.restek.com

39

GC COLUMNS |

HIGH-PERFORMANCE Rxi® COLUMNS

Rxi®-1HT, Rxi®-5HT

Rxi®-1HT Columns

(fused silica)

(nonpolar phase; dimethyl polysiloxane)

• 40% longer lifetime from specially designed fused silica tubing.

• Columns processed for high-temperature applications, such as high molecular

weight hydrocarbons.

• Temperature range: -60 to 400 °C.

Rxi®-1HT Structure

CH

3

CH

3

Si

100%

O

Similar to: (100%-methyl)-polysiloxane

similar

phases

DB-1ht, ZB-1HTinferno

15-Meter

30-Meter

ID

df

temp. limits

cat.#

price

cat.#

price

0.25 mm 0.10 µm -60 to 400 °C 13950

13951

0.25 µm -60 to 400 °C

13952

0.32 mm 0.10 µm -60 to 400 °C 13953

13954

0.25 µm -60 to 400 °C

13955

Rxi®-5HT Columns

(fused silica)

(low-polarity phase; diphenyl dimethyl polysiloxane)

• 40% longer lifetime from specially designed fused silica tubing.

• Columns processed for high-temperature applications, such as mineral oil.

• Temperature range: -60 to 400 °C.

Rxi®-5HT Structure

CH

3

CH

3

Si

95%

O

Si

5%

O

Mineral Oil on Rxi®-5HT by EN9377-2 (PTV large volume injection)

GC_EV1180

Column

Rxi®-5HT, 15 m, 0.32 mm ID, 0.10 µm

(cat.# 13906)

Sample

Mineral oil

Diluent:

Hexane

Conc.:

25 ppm

Injection

Inj. Vol.:

100 µL ptv splitless

Inlet Temp. Program: 45 °C (hold 0.45 min) to 350 °C

at 200 °C/min (hold 10 min)

Oven

Oven Temp.:

35 °C (hold 4 min) to 150 °C at 60 °C/min

to 250 °C at 50 °C/min to 350 °C at

30 °C/min

Carrier Gas

He, constant flow

Flow Rate:

2 mL/min

Detector

FID@ 360 °C

Instrument

Varian 450

Acknowledgement

Ambiente Analisi S.r.L, Italy

Sample fully eluted in just 10 min

Similar to: (5%-phenyl)-methylpolysiloxane

similar

phases

DB-5ht, VF-5ht, ZB-5HTinferno

15-Meter

30-Meter

ID

df

temp. limits

cat.#

price

cat.#

price

0.25 mm 0.10 µm -60 to 400 °C 13905

13908

0.25 µm -60 to 400 °C

13923

0.32 mm 0.10 µm -60 to 400 °C 13906

13909

0.25 µm -60 to 400 °C

13924

0.53 mm 0.15 µm -60 to 380/400 °C

13910

1

2

3

4

5 6

7 8 9

10

11

12

13

14

15

16 17

5

10

15

20

25

Time (min)

~ 2 pA bleed at 400 °C

Florida TRPH Standard Mix on Rxi®-1HT

GC_PC1272

Column

Rxi®-1HT, 15 m, 0.25 mm ID, 0.10 µm

(cat.# 13950)

Sample

Florida TRPH standard (cat.# 31266)

Diluent:

Hexane

Conc.:

50 ppm (1.25 ng on-column)

Injection

Inj. Vol.:

1.0 µL split (split ratio 40:1)

Liner:

Sky® 4.0 mm ID Precision® inlet liner

w/ wool (cat.# 23305.1)

Inj. Temp.:

275 °C

Oven

Oven Temp.:

40 °C (hold 1.0 min) to 400 °C at

20 °C/min (hold 10 min)

Carrier Gas

He, constant flow

Linear Velocity:

55 cm/sec

Detector

FID@ 400 °C

Make-up Gas

Flow Rate:

45 mL/min

Make-up Gas

Type:

N

2

Hydrogen flow:

40 mL/min

Air flow:

450 mL/min

Data Rate:

50 Hz

Instrument

Agilent/HP6890 GC

Peaks

t

R

(min)

1.

n

-Octane

1.24

2.

n

-Decane

2.71

3.

n

-Dodecane

4.19

4.

n

-Tetradecane

5.49

5.

n

-Hexadecane

6.64

6.

n

-Octadecane

7.68

7.

n

-Eicosane

8.63

8.

n

-Docosane

9.50

9.

n

-Tetracosane

10.30

10.

n

-Hexacosane

11.04

11.

n

-Octacosane

11.73

12.

n

-Triacontane

12.38

13.

n

-Dotriacontane 12.98

14.

n

-Tetratriacontane 13.55

15.

n

-Hexatriacontane 14.09

16.

n

-Octatriacontane 14.60

17.

n

-Tetracontane

15.08