286

www.restek.com

GC ACCESSORIES |

GAS MANAGEMENT

Gas Management System, Leak Detector

Restek® Gas Management System

• Removes moisture, hydrocarbons, and oxygen from carrier gas, extending

column lifetime.

• Produces high-purity carrier gas for most applications.

• Includes one each: moisture, hydrocarbon, and indicating oxygen trap.

• Replacing traps is safe and easy.

• Maximum flow: 150 cc/min for optimal performance.

Restek has put together a convenient unit providing gas purification all in one step.

Complete with an indicating trap, your gas purification issues are handled in one

central location.

The Restek® gas management system removes

water vapor (to 10 ppb), hydrocarbons (to 0.1

ppm), and oxygen (to less than 0.1 ppm) with

three traps housed in one unit.

Description

Fittings

qty.

cat.#

price

Restek Gas Management System

includes fittings for

1

/

8

" and

1

/

4

" gas line

kit

21999

Replacement Traps

qty.

cat.#

price

High-Capacity Moisture Trap

1

/

8

" Nickel-Plated Brass

ea.

21997

Capillary-Grade Hydrocarbon Trap

1

/

8

" Nickel-Plated Brass

ea.

21991

Indicating Oxygen Trap

1

/

8

" Brass

ea.

22010

22658

Verify hard-to-reach leaks using

the small probe adaptor (sold

separately).

trap configuration

did you know?

in

out

Dimensions:

12" x 14" x 3"

(30.5 x 35.6 x 7.6 cm)

Leak Detector Specifications

Detectable Gases:

Helium, nitrogen, argon, carbon dioxide, hydrogen

Battery:

Rechargeable lithium ion internal battery pack (12 hours normal operation)

Operating Temperature Range:

32–120 °F (0–48 °C)

Humidity Range:

0–97%

Warranty:

One year

Certifications:

CE, Ex, Japan

Compliance:

WEEE, RoHS

Avoid using liquid leak detectors on a GC! Liquids can be drawn into the system and/or into the leak detector.

*Caution: The Restek® electronic leak detector is designed to detect trace amounts of hydrogen in a noncombustible

environment. It is NOT designed for determining leaks in a combustible environment. A combustible gas detector should

be used for determining combustible gas leaks under any condition. When using it to detect hydrogen, the Restek®

electronic leak detector may only be used for determining trace amounts in a GC environment.

Restek® Electronic Leak Detector

Don't let a small leak turn into a costly repair—protect your analytical column by using

a Restek® leak detector.

Backed by a one-year warranty, the Restek® leak detector is the industry standard for

performance and affordability in handheld leak detectors.

Description

qty.

cat.#

price

Leak Detector With Hard-Sided Carrying Case and Universal Charger Set

(U.S., UK, European, Australian)

ea.

22655

Small Probe Adaptor for Leak Detector

ea.

22658

Dynamic Duo Combo Pack (Restek Leak Detector and ProFLOW 6000 Flowmeter)

kit

22654

Soft-Sided Storage Case for Leak Detector or ProFLOW 6000 Flowmeter

ea.

22657

Car Charger/Adaptor

ea.

22652

Universal AC Power Adaptor

ea.

22653

22655

22652

22653

Restek recommends

If you think that your Restek® electronic leak detector needs service or repair, please contact

Restek® Customer Service before sending your unit in (cat.# 22655-R). Leak detector service/repair

will include inspection and testing of the unit.