We evaluated the effectiveness of the Rt-XLSulfur
column by measuring trace sulfur
compounds in one domestic (US) and two imported brands of beer. The results from
headspace sampling of these products demonstrate the capability of the RT-XLSulfur
column and the Sulfinert
deactivated GC system to easily detect sulfur compounds at
the 20ppb level (Figure 9).
GC_FF00494
1m, 0.75mm ID Sulfinert
tubing
Rt-XLSulfur
100/120mesh (cat.# 19806)
Conc.:
headspace of a domestic (US) or imported
beer sample
Inj.:
1cc sample loop, 6-portValco
®
valve
Carrier gas:
helium
Flow rate:
10mL/ min.@ ambient temp.
Oven temp.:
60°C to 260°C@ 15°C/min. (hold 5min.)
Det. sensitivity: SCD, attn. x 1
Det. temp.:
800°C
1. hydrogen sulfide
2. methyl mercaptan
3. dimethyl sulfide and/or
ethyl mercaptan
4. unknown
Domestic Beer
ppb levels of hydrogen sulfide, dimethyl sulfide, and/or ethyl mercaptan andmethyl mercaptan in beer.
GC_FF00483
Higher amounts of dimethyl sulfide
and/or ethyl mercaptan, lower
amounts of hydrogen sulfide and
methyl mercaptan, compared to
the domestic beer.
GC_FF00487
Equivalent amounts of dimethyl sul-
fide and/or ethyl mercaptan, lower
amounts of hydrogen sulfide and
methyl mercaptan, compared to the
domestic beer.
Imported Beer (Mexico)
Imported Beer (Canada)
Figure 9