CryoSulfur GC System
- DISCONTINUED
6
|
|
|
|
60 meter Capillary Column
10-port Gas Sampling Valve
Combination FPD/FID Detectors
CryoCooled Peltier Trap Accessory
Vacuum Pump and Interface for Reproducible Sampling
Built-in, "whisper quiet" Air Compressor
4 channel PeakSimple Data System
...on the compact
chassis
The SRI CryoSulfur GC comes with everything you need to detect low-level sulfur compounds in gas samples. Since
some sulfur compounds do not trap well, the CryoSulfur GC uses the CryoCooled Peltier Trap Accessory to enrich
the sample, providing lower detection limits. Like all SRI traps, the CryoCooled Peltier Trap Accessory is plumbed as the
loop of a 10-port gas sampling valve. It can cool down to -15
o
C for sample enrichment. After enrichment, the CryoCooler
is heated (up to 200
o
C), and the valve injects the sample into the 60-meter capillary column. Once the sample
components are separated by the column, they are detected by the Flame Photometric and Flame Ionization detectors.
The CryoSulfur GC uses a vacuum pump (included) to draw gas samples into the
CryoCooler. You can sample ambient air, or use the provided adaptor to connect a
Tedlar bag. The vacuum pump interface, which is an electrical outlet on the left-hand
sife of the GC, allows the vacuum pump to be turned ON/OFF by the PeakSimple data
sysrem to provide consistent sampling times.
This chromatogram, generated by an
SRI CryoSulfur GC, shows the FPD
response to 10ppb hydrogen sulfide
(H
2
S) as enriched by the CryoCooler
at -10
o
C.
Sample: 10ppb H
2
S
Column: 60 meter MXT-1
Carrier: helium at 10mL/minute
Vacuum pump: 20mL/minute
Trap: -10
o
C
FPD gain: HIGH
FPD temperature: 150
o
C
FPD volts: 500