Page 118 - SRI012_1-152

Basic HTML Version

110-Vial Liquid Autosampler - SRI Instruments
HT300A 110-Vial Liquid Autosampler
DISCONTINUED
83
|
|
|
|
Holds 110 2mL or 2.5mL vials
Interfaces with SRI and other GCs
15-Step Automatic Injection Sequence
Direct Injection, No Transfer Lines
The HT300A Liquid Autosampler is made to meet the high
throughput liquid injection needs of your GC analysis. Like the
HT200H Headspace Autosampler, the swivel head design
simulates the movements of manual direct injection and eliminates
the need for transfer lines, as well as leaving the injection port free
for manual injections. Up to 10 analytical methods, including
function speeds, may be stored in the HT300A's memory.
The automatic injection sequence may have up to 15
steps, which may be programmed to include:
O
PERATING
S
PECIFICATIONS
Sampling
Sample volume
Air volume
Aspirating speed
Needle washing
Air bubble removal
Viscosity time
Syringe sizes
Injection
Injection speed
Waiting time before & after injection
Injection depth
Internal Standard Technique
IS volume
Air gap volume
Mode
Sample
steps of 0.1uL
steps of 0.1uL
0.1uL - 100uL/second
up to 15
up to 15 strokes
1-60 seconds
1, 10, 25, 50, 100uL (1000uL
large volume version)
Injection
0.1uL - 100uL/second
1-60 seconds
variable
IST
steps of 0.1uL
steps of 0.1uL
1 or 2 air gaps
First sample of group
Last sample of group
Analytical method
Number of injections for each sample
Pre and post washing solvent position
Internal standard position (if used)
The sampling system eliminates air bubbles, and the
variable fill speed allows for a wide range of sample
viscosities. The syringe may be washed with solvent or
sample.
8690-3000
110-Vial Liquid Autosampler
DISCONTINUED
HT300A 110-Vial Liquid Autosampler
83
|
|
|
|