Page 33 - PermeationTubes-Metronics11_1-60

This is a SEO version of PermeationTubes-Metronics11_1-60. Click here to view full version

« Previous Page Table of Contents Next Page »
Specifications
24 cm (9.5")
1.6 cm (0.62")
Permeation device
Maximum length
Maximum diameter
Permeation chamber
Teflon
®
or Pyrex
®
Will accomodate multiple devices
Flowmeters
High resolution 15 cm graduated scale.
Certified ±1% accuracy full scale, NIST
traceable. Full scale ranges of 1,2,5,10,
15, and 20 L/min.
Flowmeter calibration accuracy
( max % deviation from float setting)
At maximum flow
At minimum flow
1 %
3 %
Ambient operating temperature
10°C - 40°C
Temperature indicator
Digital display
5°C above ambient to 110°C
Temperature control range
Permeation chamber temperature
equilibration time
< 1 hr
Pumps
Diaphragm pumps for continuous operation.
Available for 10 or 20 L/min capacity
External gas stream supply
and modification
Instruments with separate stream inlets or
stream bypasses have provision for
connecting dehydrators, special scrubbers,
filters, etc. in series with the carrier and/or
dilution gas streams
Case dimensions
17" wide x 23.75" deep x 10.5" high
(43.2 cm x 60.3 cm x 26.7 cm)
Instrument weight
Units with internal pumps
Model 230
Model 340
Model 450
Units without internal pumps
Model 230
Model 340
Model 450
55 lbs. (25.0 kg)
56 lbs. (25.4 kg)
59 lbs. (26.8 kg)
43 lbs. (19.5 kg)
44 lbs. (20.0 kg)
47 lbs. (21.3 kg)
Power consumption
Units with internal pumps
Model 230
Model 340
Model 450
Units without internal pumps
Model 230
Model 340
Model 450
290W
300W
340W
100W
110W
145W
Instrument noise emission (at 3 ft)
Units with internal pumps
Units without internal pumps
55 - 60 dBa
45 - 50 dBa
US customers
call toll-free
(877) 737-1887
VICI
®
is a registered trademark of Valco Instruments Co. Inc. and VICI AG. Dynacalibrator
®
and Dynacal
®
are registered trademark of VICI Metronics, Inc
Teflon
®
is a registered trademark of E.I. duPont de Nemours. Pyrex
®
is a registered trademark of Corning Glass.
North America, South America, and Australia/Oceania contact:
Europe, Asia, and Africa contact:
Temperature controller
Solid state proportional with 0.1°C set point
accuracy, NIST traceable.
Metronics Inc.
tel: 360 697-9199
fax: 360 697-6682
metronics@vici.com
VICI AG International
rev 4/06
PB-026 dynacal
tel: Int + 41 41 925-6200
fax: Int + 41 41 925-6201
info@vici.ch
Website : www.chromtech.net.au E-mail : info@chromtech.net.au TelNo : 03 9762 2034 . . . in AUSTRALIA