FUME HOOD
BSC15IIIX/ZX/KX, Class III
Model
BSC15IIIX
BSC15IIIZX
BSC15IIIKX
External Dimensions (mm)
L1340xW850xH2250
L1645xW850xH2250
L1950xW850xH2250
Internal Dimension (mm)
L1240xW650xH650
L1545xW650xH650
L1850xW650xH650
No. of gloves
2
4
4
Delivery window dimension(mm)
L480xW400xH400
Interception efficiency
99.999% (for 0.12µm particulates)
Noise level
<60dBA
Complete sealed
Front window is sealed, and with pairs of gloves connecting to the operation
area, product is transferred by the delivery window. And there is no out-
connected valve.
100% exhaust
100% air exhaust, no air recirculation, and air is exhausted by three ULPA filters
and to the room directly. If the air needs to be exhausted to outdoors, another
out connected duct need to be ordered, and extra duct and motor are required.
Complete negative pressure
The two motors design makes interior of cabinet is complete negative pressure,
and can reach -274pa, triple layers filter film, ensures operation’s security,
neither substance nor polluted air will leak out.
Model
BSC7IIA2
External size (mm)
700x500x1200 (WxDxH)
Internal size (mm)
600x400x580 (WxDxH)
Material
Work zone interior is made of 304 stainless steel
expoxy coated cold rolled steel
Control system
Microprocessor
Display
LED display
Main filter
Hepa filter with 99.999% efficiency at 0.3 micros
Exhaust filter
Hepa filter with 99.999% efficiency at 0.3 micros
Intake airflow velocity
0.53m/s
Exhaust airflow velocity
0.33m/s
Air circulation
70%
Air exhaust
30%
Front sash
Sliding by electronic control, double layer
toughened glass
Clean level
Class 100
Noise level
≤65db
Lightning
≥680Lux
UV lamp
15W x 1
Power
220V, 50Hz ±10%
Compliance
CE, ISO13485, ISO9001, ISO14001
BSC7IIA2, Class II
The Class II Biological Safety Cabinet is designed with inward air flow at a
velocity to protect personnel, HEPA-filtered vertical laminar airflow for product
protection, and HEPA-filtered exhausted air for environmental protection.
Biological Safety Cabinet
BSC7IIA2
BSC15IIIX