Page 5 - MRC11_1-292

This is a SEO version of MRC11_1-292. Click here to view full version

« Previous Page Table of Contents Next Page »
6
A
ANALYZERS
Semi/Fully-Automated Chemistry
Features:

Easy keypad operation with LCD display

Continuously adjustable grating,
enhances the stability and reliability of
result.

Reagent open system supports !ow
cell and cuvette mode, close system on
request.
Model
SACA-19900
Principle
Absorbance
Absorbance Range
-0.500 ~ 3.500Abs
Resolution
0.001 Abs (displayed), 0.0001Abs (calculated)
Light Source
Halogen lamp
Wavelengths
Continuously adjusted grating, wavelength range is
330 - 800nm, 1nm by step adjustment
Wavelength Accuracy
± 1nm
Band Width
< 6nm
Flow cell
Metal - quartz !ow cell
Temperature Control
25°C,30°C,37°C,±0.1°C and ambient temperature
Aspirate Volume
0~3000µl
Carryover
< 1%
Memory
Up to 204 protocols and 5400 results can be stored
Display
LCD display
Interface
RS-232
Output
Built-in printer, or external printer (optional)
Power Supply
AC 110V - 220V ± 10%, 50-60Hz
Net Weight
8KG
Dimensions (mm)
L360xW318xH185
SACA-19900
SACA-19900, Semi-Auto Chemistry, Grating Inside, Easy Keypad Operation With LCD Display

Analytical modes include kinetic, "xed time, bichromatic end point with or without reagent blank or sample blank, linear or
non-linear calibration curves

Powerful QC program: Westguard multi-rule, Levey-Jennings plots,2 controls per test

Large
memory, up to 204 protocols and 5400 results can be stored

Additional recloser design with lamp saving

Powerful internet
connection enables remote data sharing & OS update

Compact design enables USER-DIY maintenance

Multi-language
software available on request.
FACA-1240,Fully Automated Chemistry, Discrete, Constant Throughput 120 Test/Hours, RandomAccess
Features:

Automatic, Random Access.

Up to 240 tests per hour.

Reagent pre-heating, liquid level
detection.

Automatic 10 steps washing system
(optional), Low carry-over.

Labor saving. Simple programming and
real walk-away operation.

24 hours non-stop Reagent cooling
compartment.

Reagent open system, close system on
request.

Micro-volume for sample and reagent.

User-friendly software based on
Windows XP.
FACA-1240
Model
FACA-1240
Throughput
200 tests/hour, Constant throughput around 120 tests/hour
Contents of the testing
Clinical chemistry, Immunoturbidimetric
Analytical modes
End-point, Two-point, Kinetic
Programming
Open system with user de"ned programs and calculation
Sample/Reagent tray
40 positions for samples and 40 positions for
reagents in refrigerated compartment
Refrigerated temp.
4 ~ 15°C
Reagent volume
R1: 180 ~ 450 #L, 1 #L adjustable
R2: 30 ~ 250 #L, 1 #L adjustable
Sample volume
3 ~ 45 #L, 0.5 #L adjustable
Sample & Reagent probe
Collision protection,Liquid level detection & Inventory checking
Probe cleaning
Automatic washing both inside and outside, carry over <0.1%
Reaction positions
81 cuvettes, 9 cuvettes/strip
Cuvette
Optical length 5mm
Reaction volume
180 ~ 500ul
Reaction temp.
37°C, !uctuation ±0.1°C
Lamp
Halogen lamp
Absorbtion range
0 ~ 3.500Abs±1%
Wavelengths
8 wavelengths, including 340nm,405nm,450nm,510nm,
546nm,578nm,630nm,670nm
Resolution
0.0001ABS
Power supply
AC 100/240V 50/60 Hz
Temperature
10 ~ 35°C
Humidity
35 ~ 80%
Water consumption
Max 2.5 L/hour
Dimention(cm)
L800xW600xH650
Gross weight
80KG
Automatic, Randam Access, STAT sampling priority, Barcode-Reading support (optional)
Pre-heating for reagent
Sample/Reagent Handling
System Function
Reaction System
Measuring and Optic System
Working Condition
Independent mixing probe, Automatic washing system
SACA-19900
SACA-19900
FACA-1240