Page 243 - MRC11_1-292

This is a SEO version of MRC11_1-292. Click here to view full version

« Previous Page Table of Contents Next Page »
298
V
VISCOSITY
Rotary Viscometers
VIS-S3
Direct readout of all measurement parameters on an illuminated Liquid Crystal
Display
Data on screen:
Speed selected: Lp.m.
Spindle selected: S.P.
Viscosity reading: cP (mPa-s)
Percentage of full scale: %
Sample
temperature: °C
Temperature determinated by PT100 probe
Viscosity
reading: dynamic viscosity (cP or mPa-s)
The viscometer and the PT100 can
be calibrated by the user
Auto range function to determine the maximum
viscosity with each spindle/speed combination
It can work with a micro printer
Ultra-sensitive front keyboard that allows easy input of all parameters
"ERROR" message in the screen with sound alarm when the viscosity of the test
material exceeds the full scale of the selected combination (SP/r.p.m.)
Sound
alarm if the equipment is working at under 20% of selected full scale
The
viscometer is equipped with many pre-sets speeds, from 0.3 to 100 r.p.m
Safety protection against sudden rises in the power supply
English language
options
Direct reading in cP (mPa-s).
VIS-8, Digital Basic Rotary Viscometers
Accuracy
± 1% of the full scale
Resolution
Using LCP (Low Viscosity Adapter): 0.01,
When viscosity is
lower than 10,000 cP:1, When viscosity is equal or higher than
10,000 cP:1
Repeatability
0.5%
Temperature Range
from 0.0
o
C to + 100.0
o
C
Resolution
0.1
o
C
Accuracy
±0.1
o
C
Type of Probe
PT 100
Outputs
RS232, Micro Printer
Power
All Models are Supplied with 85/265 VAC
Net Weight
4 kg
Accuracy
± 1% of the full scale
Resolution
Using LCP (Low Viscosity Adapter): 0.01, When viscosity is
lower than 10,000 cP:1, When viscosity is equal or higher than
10,000 cP:1
Repeatability
0.5%
Temperature Range
from 0.0
o
C to + 100.0
o
C
Resolution
0.1
o
C
Accuracy
±0.1
o
C
Type of Probe
PT 100
Outputs
Computer interface RS232, Micro Printer
Power
All Models are Supplied with 85/265 VAC
Net Weight
4 kg
Model
Measuring Range
Fixed Speeds in R.P.M
VIS-8
10-2,000,000 Cp
0.3,0.6,1.5,3,6,12,30,60
Model
Measuring Range
Fixed Speeds in R.P.M
VIS-S1
10-600,000 cP
1-60(nonpolar shift)
VIS-S2
10-6,000,000 cP
0.1-99.9
VIS-S3
10-80,000,000 cP
0.1-99.9
VIS-AI
10-1,000,000 cP
0.1-200.0
VIS-S1/S2/S3/AI, Digital Rotary Viscometers, Without Limits
Direct readout of all measurement parameters on an illuminated Liquid Crystal Display
Data on screen:
Speed selected:
r.p.m.
Spindle selected: S.P.
Viscosity reading: cP (mPa-s)
Percentage of full scale: %
Sample temperature: °C
Temperature deterrninated by PT100 probe
Viscosity reading: dynamic viscosity (cP or mPa-s)
The viscometer and the
PT100 can be calibrated by the user
Auto range function to determine the maximum viscosity with each spindle/speed
combination
It can work with a micro printer
Ultra-sensitive front keyboard
that allows easy input of all parameters
“ERROR” message in the screen with
sound alarm when the viscosity of the test material exceeds the full scale of the
selected combination (Slvr.p.m.)
Sound alarm if the equipment is working at
under 20% of selected full scale
The viscometer is equipped with many
pre-sets speeds, from 0.1 to 100 r.p.m
Safety protection against sudden rises
in the power supply
English language options
Direct reading in cP (mPa-s).
VIS-8