Page 139 - MRC11_1-292

This is a SEO version of MRC11_1-292. Click here to view full version

« Previous Page Table of Contents Next Page »
179
M
METERS-WATER QUALITY
TU-2016
TU-2016, Turbidity Meter
TU-2016
Model
Custom one-chip of microprocessor LSI circuit.
Circuit
LCD size: 41mm x 34mm
Display
0.00 to 50.00NTU, 50 to 1,000NTU
NTU : Nephelometric Turbidity Unit
Auto range
Range
± 5 % F.S. or ± 0.5 NTU, which ever is greater
Accuracy
LED, 850nm
Light source
Photo diode
Detector
Meet ISO 7027.
Standard
Less than 10 seconds
Response time
10 mL
Sample volume needed
Freeze the display reading
Data Hold
Maximum & Minimum value
Memory Recall
Approx. 1 second
Display Sampling Time
Auto shut off saves battery life or manual off by push button.
Power off
0NTU, 100NTU
Calibration points
0 to 50
O
C
Operating Temperature
Less than 85% R.H.
Operating Humidity
DC 1.5 V battery ( UM4, AAA ) x 6 PCs, or equivalent
Power Supply
Stand by: Approx. DC 3.5 mA.
Power Current
Testing: Approx. DC 36 mA
320 g/0.70 LB. @ Battery is included.
Weight
155x76x62mm (6.1x3.0x2.4inch)
Dimension
Instruction manual - 1 PC Included
Testing bottle with 0 NTU standard solution - 1 PC
Testing
bottle with 100 NTU standard solution....1 PC
Empty testing bottle - 2 PCs
Clean cloth - 1 PC
Clean solution (Distill water) - 1 bottle
Hard carrying case, CA-08 - 1 PC.
Accessories Included
Testing bottle with 100 NTU standard solution, TU-100NTU
Empty testing bottle, 0601.
Optional Accessories
Features:
Designed to meet ISO 7027.
NTU (Nephelometric TURBIDITY Unit) measuring unit.
Wide and auto measurement range : 0 to 1,000 NTU.
High resolution : 0.01 NTU/1 NTU.
The unique optics structure, enables the instrument to read low value of
TURBIDITY to the high level up to 1,000 NTU.
Four operation buttons and two calibration points, easy operation and
giarantee the spec. accuracy.
Jumbo LCD, easy readout.
Microprocessor circuit assures maximum possible accuracy, provides special
functions and features.
Battery operated for field and on-site testing convenience.
Data hold function for freezing the desired value on display.
Records Maximum and Minimum readings with Recall.
Heavy duty & compact housing with hard carrying case, designed for easy
carry out & operation.
Auto shut off is available to save battery life.
Application : Test municipal water, food and beverage water, or other aqueous
solution where fluid clarity is important.
Turbidity