Model
HC30R-2
HC9R-2
Capacity
865 Liter
283 Liter
Interior dimensions (cm)
W78.7xD69.8xH157.5
W76.2xD50.8xH76.2
Exterior dimensions (cm)
W109xD89xH216
W111.76xD73.66xH152.4
Temperature range
10°C to 70°C
10°C to 70°C
Temperature uniformity
±0.5°C at 37°C
±0.5°C at 37°C
Electrical specification
Volts: 220V Hertz: 50/60 Hz Watts:
1650W Amps: 7.5A
Volts: 220V Hertz: 50/60 Hz Watts:
1430W Amps: 6.5A
Jacket type
Air
Air
Relative humidity range
Ambient +10% to 95%
Ambient +10% to -95%
Shelving
6 Supplied (16 Maximum)
3 Supplied (8 Maximum)
HC9R-2
HC30R-2
MRC Humidity Test Chambers provide a controlled
environment for a wide range of industrial and
biotechnical testing applications.
This line is designed to duplicate natural
conditions, that allows you to determine the
limitations of a sample when exposed to various
temperature and moisture fluctuations.
This unit incorporates a refrigeration system that dramatically
increases the operational range of the cabinet. A low-pressure
water vapor generator, injecting saturated water vapor into the
recirculating air duct, controls chamber humidification.
This process is preferable to steam generation because steam
introduces additional heat to the chamber atmosphere, which
then compromises temperature control. Uniform temperature
and humidity are maintained by circulating chamber
atmosphere through a recirculating air duct system. The air
ducts also serve as shelf supports and can be easily
removed for cleaning.
Microprocessor Controls and Digital Display for both Temp.
and Humidity. Microprocessor temperature and humidity
controls with digital display provides optimal stability and allow
for accurate control. Controllers have easy-to-use keypad,
touch pads and a calibration feature. Setpoints have memory
capabilities in the event of power outages.
Safety Features.
The MRC safety controller provides
independent, overtemperature protection (OTP) that operates
independently from the main controller. The dual-control,
Features/ Benefits:
Stainless steel chambers are seamless for durability, corrosion resistance and quality
Low-pressure
water vapor generator system provides humidification without generating additional heat
Microprocessor controls and digital
readout display both temperature and humidity values
Refrigerated models provide extended performance for temperature
and RH ranges
Two layers of insulation to maintain temperature and reduce heat loss.
Applications:
Shelf Life and Half-Life Testing
Packaging
Sterility Testing
Mil-Specifications
Component Burn-In
Vapor Transmission.
fail-safe
function means
a second
system will
ensure that
temperature
setpoint is
never
exceeded.
By operating
independently,
it will take over
& control the
heating
function if the
main temp.
setpoint is
exceeded.